Acta Photonica Sinica, Volume. 49, Issue 8, 0822002(2020)

Optical System Design of Miniaturized Fundus Camera Based on Non-coaxial Array Illumination

Yan-xiong WU... Zhi-yuan GUO, Miao YU and Li-ping WANG* |Show fewer author(s)
Author Affiliations
  • Foshan University, Foshan, Guangdong 528000, China
  • show less
    Figures & Tables(20)
    Diagram of light path structure of traditional fundus camera
    Diagram of light path structure of fundus camera based on non-coaxial ring array illumination
    Analysis diagram of effective illumination area of fundus
    Analysis diagram of fundus lighting area
    Effective imaging FOV analysis chart
    Result diagram of optical system design
    Result diagram of lighting sub array system design
    Result diagram of circular array lighting system design
    Analysis diagram of lighting uniformity simulation
    Diagram of simulation of fundus illumination effect
    Simulation of stray light in new fundus camera
    Simulation of stray light in traditional fundus camera
    Two dimensional structure diagram of imaging system
    Curve diagram of MTF of imaging system
    Spot diagram and distortion map of imaging system
    • Table 1. Design parameters of lighting system

      View table
      View in Article

      Table 1. Design parameters of lighting system

      ParameterValue
      Wave band/nm935~945
      Divergence angle of light source /(°)30
      Pupil spot diameter range /mm4~6
      Working distance /mm10
      Fundus illumination height /mm12
      Tocal length of system/mm≤20
    • Table 2. Multiple structural parameters of lighting subarray system

      View table
      View in Article

      Table 2. Multiple structural parameters of lighting subarray system

      OperandConfig1Config2Config3Config4Config5Config6
      CATX-15-15-15-15-15-15
      CADY-7.5-7.5-7.5-7.5-7.5-7.5
      CATZ060120180240300
    • Table 3. Simulation data of fundus illumination uniformity

      View table
      View in Article

      Table 3. Simulation data of fundus illumination uniformity

      Num sampTotal power/WMax value/(W·mm-2)Min value/(W·mm-2)Average/(W·mm-2)
      Receiver1 003 21610.2470.086 0730.000 00.050 859
    • Table 4. Design parameters of imaging system

      View table
      View in Article

      Table 4. Design parameters of imaging system

      ParameterValue
      Wave band /nm935~945
      Line field of view /mm±5.24
      Object resolution /μm6
      Pupil diameter/mm2
      Total length of system/mm≤130
      Distortion /%< 5
      MTF value> 0.2@91 lp/mm
    • Table 5. Lens data of imaging system

      View table
      View in Article

      Table 5. Lens data of imaging system

      No.Surface typeRadius/mmThickness/mmMaterialRefractive indexAbbe number
      OBJStandard12.500.00
      1Standard12.5017.251.3455.00
      2Even Asphere5.674.001.4255.00
      3Standard-10.480.00
      STOStandard-10.103.001.3455.00
      5Standard-3.630.601.3855.00
      6Standard-6.8929.99
      7Standard64.705.22H⁃K9L1.5264.21
      8Standard-32.011.00
      9Standard20.735.41H⁃K9L1.5264.21
      10Standard214.4331.60
      11Standard27.251.92H⁃LAK31.7551.00
      12Standard-15.694.15
      13Standard-3.851.70H⁃F21.6137.00
      14Standard19.691.00
      15Standard-17.192.11H⁃LAK31.7551.00
      16Standard-5.731.00
      17Standard42.392.06H⁃LAK31.7551.00
      18Standard-12.9618.00
      19StandardInfinity-0.07
      IMAStandardInfinity0.00
    Tools

    Get Citation

    Copy Citation Text

    Yan-xiong WU, Zhi-yuan GUO, Miao YU, Li-ping WANG. Optical System Design of Miniaturized Fundus Camera Based on Non-coaxial Array Illumination[J]. Acta Photonica Sinica, 2020, 49(8): 0822002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Feb. 28, 2020

    Accepted: May. 19, 2020

    Published Online: Nov. 27, 2020

    The Author Email: WANG Li-ping (home313@126.com)

    DOI:10.3788/gzxb20204908.0822002

    Topics