Optics and Precision Engineering, Volume. 30, Issue 1, 31(2022)

Calibration of polarization axis direction of sub focal plane polarization image sensor

Yang BAI1... Kaichun ZHAO1,2,*, and Zheng YOU12 |Show fewer author(s)
Author Affiliations
  • 1Department of Precision Instrument, Tsinghua University, Beijing00084, China
  • 2State Key Laboratory of Precision Measurement Technology and Instrument, Tsinghua University, Beijing100084, China
  • show less
    Figures & Tables(10)
    Structure of pixel array of sub focal plane polarization image sensor
    Structure of calibration device for sub focal plane polarization image sensor
    Principle of calibration method for sub focal plane polarization image sensor
    Simulation sample points of single pixel
    Calibration devices for polarization axis direction of sub focal plane polarization image sensor
    Fitting results of response values of each polarization channel changing with angle
    • Table 1. Results of polarization axis direction simulation calibration

      View table
      View in Article

      Table 1. Results of polarization axis direction simulation calibration

      Number04590135
      10.0044.9990.00135.00
      20.0044.9990.00135.00
      30.0044.9990.00135.00
      40.0044.9990.00135.00
    • Table 2. Composition and equipment parameters of calibration devices

      View table
      View in Article

      Table 2. Composition and equipment parameters of calibration devices

      NameParameterValue
      Integrating sphereUniformity>98%
      PolarizerExtinction ratio>1000
      TurntableMin step size1/800°
    • Table 3. Fourier fitting result of sampling sequence

      View table
      View in Article

      Table 3. Fourier fitting result of sampling sequence

      Nominal value/(°)a0a1b0wR2
      0165.4053.81-95.6420.999 6
      45169.2095.6455.2720.999 4
      90170.40-55.3095.6120.999 6
      135166.50-96.33-55.3120.999 4
    • Table 4. Nominal and actual values of polarization principal axis direction

      View table
      View in Article

      Table 4. Nominal and actual values of polarization principal axis direction

      Nominal valueCalibration valueDifference
      00.0000
      4545.0900.090
      9090.1000.100
      135135.0100.010
    Tools

    Get Citation

    Copy Citation Text

    Yang BAI, Kaichun ZHAO, Zheng YOU. Calibration of polarization axis direction of sub focal plane polarization image sensor[J]. Optics and Precision Engineering, 2022, 30(1): 31

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Modern Applied Optics

    Received: May. 11, 2021

    Accepted: --

    Published Online: Jan. 20, 2022

    The Author Email: ZHAO Kaichun (kaichunz@mail.tsinghua.edu.cn)

    DOI:10.37188/OPE.20223001.0031

    Topics