Infrared and Laser Engineering, Volume. 50, Issue 8, 20200471(2021)

Hybrid PSO-AMLS-based method for data fitting in the calibration of the infrared radiometer

Sen Yang
Author Affiliations
  • College of Mechanical and Electrical Engineering, Northeast Forestry University, Harbin 150000, China
  • show less
    Figures & Tables(12)
    Flowchart of the hybrid PSO-AMLS-based method
    The fitting errors δ of the LS and PSO-AMLS method when the independent variables are uniform distribution: (a) nnodes=11; (b) nnodes=21
    The fitting errors δ of the ALS and the PSO-AMLS method when the independent variables are non-uniform distribution
    Diagram of the infrared radiometer
    The fitting errors δ of the LS and ALS method: (a) LS method; (b) ALS method
    The fitting errors of the PSO-AMLS and ALS method
    The deviation δ* obtained for different fitting methods
    • Table 1. The data used in the first example

      View table
      View in Article

      Table 1. The data used in the first example

      xyxyxy
      0173.6458144.7417
      1283.8284154.873
      22.414294165
      32.7321104.1623175.1231
      43114.3166185.2426
      53.2361124.4641195.3589
      63.4495134.6056205.4721
    • Table 2. The data used in the second example

      View table
      View in Article

      Table 2. The data used in the second example

      yxyxyx
      11611.51146.5
      21.5716.51256.5
      32.5822.51367.5
      44.5929.51479.5
      57.51037.51592.5
    • Table 3. Main specifications of the infrared radiometer

      View table
      View in Article

      Table 3. Main specifications of the infrared radiometer

      ComponentParameter
      Optical lensDiameter: 70 mm
      Field of view: ±0.6°
      Spectral channelsMW: 3.7-4.8 µm
      DetectorType: Insb
      A/DPrecision: 16
      Frequency: 100 kHz
      Preamplifier106×
    • Table 4. Main specifications of the blackbody-collimator

      View table
      View in Article

      Table 4. Main specifications of the blackbody-collimator

      ComponentParameter
      BlackbodyModel: HFY-200C
      Temperature: 5-1000 ℃
      ApertureDiameter: 0.1-13.5 mm
      CollimatorModel: HGD-1
      Focal length: 650 mm
    • Table 5. The calibration data of T and V in the calibration of the infrared radiometer

      View table
      View in Article

      Table 5. The calibration data of T and V in the calibration of the infrared radiometer

      V /V T /℃ V /V T /℃
      0.021001.354450
      0.051501.828500
      0.1172002.383550
      0.2272502.994600
      0.3943003.75650
      0.6323504.555700
      0.949400
    Tools

    Get Citation

    Copy Citation Text

    Sen Yang. Hybrid PSO-AMLS-based method for data fitting in the calibration of the infrared radiometer[J]. Infrared and Laser Engineering, 2021, 50(8): 20200471

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Infrared technology and application

    Received: Dec. 7, 2020

    Accepted: --

    Published Online: Nov. 2, 2021

    The Author Email:

    DOI:10.3788/IRLA20200471