Journal of Radiation Research and Radiation Processing, Volume. 42, Issue 6, 060209(2024)

Investigating the performance of polytetrafluoroethylene solid free radical dosimeter

Jingfeng CHANG1... Guanghua ZHANG1, Guozhong WU2, Rongfang SHEN2, Kaifeng DONG1 and Kanghai ZHENG1,* |Show fewer author(s)
Author Affiliations
  • 1Shanghai Jinpengyuan Irradiation Technology Co.,Ltd.,Shanghai 201707,China
  • 2Shanghai Institute of Applied Physics,Chinese Academy of Sciences,Shanghai 201800,China
  • show less
    Figures & Tables(11)
    Hpp signals of different dose values
    Response curve of PTFE solid dosimeter tablets (background subtracted) (a) and partial enlarged drawing (b)
    ESR response vs. absorbed dose curves of PTFE dosimeter
    Gamma irradiation field dose rate distribution map
    Dose response at different measurement times after irradiation
    Responses curves of PTFE to different rays
    • Table 1. Zero dose ESR response test data

      View table
      View in Article

      Table 1. Zero dose ESR response test data

      样品

      Sample

      重量/ mg

      Weight

      标准物质信号 / a.u.

      Standard material signal

      剂量计信号Hpp / a.u.

      Dosimeter signal

      响应 / mg-1

      Response

      信噪比 / dB

      Signal-to-noise ratio

      腔内温度 / ℃

      Intra-cavity temperature

      163.005 914.6118.290.2950.8432.25
      263.805 901.0325.170.3970.8632.21
      363.605 912.0919.360.3151.6932.19
      463.605 903.8123.540.3751.7032.15
      563.905 915.3131.950.5047.6632.12
      663.405 905.6429.420.4743.8932.10
      763.605 901.4127.120.4259.5732.08
      863.905 910.1327.180.4372.4132.06
      963.205 901.4119.790.3155.7032.16
      1063.005 899.5127.580.4476.64531.89

      均值

      Average

      63.50N/A24.940.3958.10N/A
    • Table 2. PTFE dosimeter flipping uncertainty test data table

      View table
      View in Article

      Table 2. PTFE dosimeter flipping uncertainty test data table

      样品

      Sample

      旋转角度

      Rotation angle

      剂量计信号Hpp / a.u.

      Dosimeter signal

      响应 / mg-1

      Response

      响应均值d¯i / mg-1

      Average response intensity

      标准偏差SDi

      Standard deviation

      125 457.52382.41381.980.38
      90°25 441.99381.69
      180°25 443.68381.64
      270°25 448.05382.18
      226 143.66392.68392.680.67
      90°26 115.58392.24
      180°26 135.29392.19
      270°26 233.71393.62
      326 415.28396.42397.051.16
      90°26 435.03396.57
      180°26 414.78396.43
      270°26 532.23398.78
      426 090.45391.44392.340.97
      90°26 100.86391.75
      180°26 129.13392.57
      270°26 194.38393.61

      变异系数 Coefficient of variation[14]

      CV%=(ni-1)(SD2/d¯i2)1/2/(ni-1)×100%

      0.06%
    • Table 3. Uncertainty test data for PTFE dosimeter flip

      View table
      View in Article

      Table 3. Uncertainty test data for PTFE dosimeter flip

      样品

      Sample

      翻转前后

      Flip top to bottom

      剂量计信号Hpp/ a.u.

      Dosimeter signal

      响应 / mg-1

      Response

      响应均值d¯i / mg-1

      Average response intensity

      标准偏差SDi

      Standard deviation

      1之前Before26 539.26398.94398.051.25
      之后After26 481.60397.17
      2之前Before26 158.24392.86393.490.89
      之后After26 193.12394.11
      3之前Before26 199.79393.77394.511.04
      之后After26 268.01395.24
      4之前Before25 441.96382.15382.290.20
      之后After25 447.03382.44

      变异系数 Coefficient of variation[14]

      CV%=(ni-1)(SD2/d¯i2)1/2/(ni-1)×100%

      0.12%
    • Table 4. Comparison of irradiation responses at different dose rates

      View table
      View in Article

      Table 4. Comparison of irradiation responses at different dose rates

      辐照点位

      Radiation point position

      剂量计信号Hpp / a.u.

      Dosimeter signal

      吸收剂量 / kGy

      Absorbed dose

      平均剂量 / kGy

      Average dose

      偏差[(A-B)/B] /%

      Deviation [(A-B)/B]

      静态吸收剂量率范围3.0~4.0 kGy/h

      Static absorbed dose rate range

      10 053.1514.00

      A

      13.94

      0.61
      10 043.5513.96
      9 929.9613.59
      10 124.6814.22

      动态吸收剂量率范围2.7~122.2 kGy/h

      Dynamic absorbed dose rate range

      10 106.3714.10

      B

      13.86

      10 047.3513.93
      10 001.6713.81
      9 936.7813.59
    • Table 5. Uncertainty analysis of PTFE dosimeter

      View table
      View in Article

      Table 5. Uncertainty analysis of PTFE dosimeter

      不确定度分量Uncertainty componentsA / %B / %
      校准曲线拟合Calibration curve fitting2.12
      剂量计重复性PTFE dosimeter repeatability PTFE0.20
      溯源不确定度Traceability uncertainty1.15

      参考标准剂量计和校准剂量计在剂量上的差异

      Dose difference between reference standard dosimeter and calibration dosimeter

      0.34
      红宝石信号修正Ruby signal correction0.27
      剂量计重量称量偏差Dosimeter weight measurement deviation0.45
      旋转影响Dosimeter weight measurement deviation0.06
      翻转影响Flip effect0.12
      开始温度修正Initial temperature correction0.10
      结束温度修正Final temperature correction0.12

      方和根法合并A类/B类不确定度

      The square and root methods combine Class A/Class B uncertainty

      2.131.32
      合成不确定Combined uncertainty2.51
      扩展不确定度(k=2) Expanded uncertainty5.02
    Tools

    Get Citation

    Copy Citation Text

    Jingfeng CHANG, Guanghua ZHANG, Guozhong WU, Rongfang SHEN, Kaifeng DONG, Kanghai ZHENG. Investigating the performance of polytetrafluoroethylene solid free radical dosimeter[J]. Journal of Radiation Research and Radiation Processing, 2024, 42(6): 060209

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: RADIATION CHEMISTRY

    Received: Nov. 19, 2024

    Accepted: Dec. 19, 2024

    Published Online: Jan. 15, 2025

    The Author Email: ZHENG Kanghai (郑康海)

    DOI:10.11889/j.1000-3436.2024-0106

    Topics