Laser & Optoelectronics Progress, Volume. 58, Issue 23, 2330001(2021)

Performance Comparison of Pulsed and Continuous Quantum Cascade Lasers in Nitric Oxide Detection

Zhenhan Li1,2、*, Yue Yu3, and Rui Guo4
Author Affiliations
  • 1School of Automation, Guangdong University of Technology, Guangzhou , Guangdong 510006, China
  • 2Beijing Reyin Instrument Technology Co., Ltd., Beijing 100088, China
  • 3China Special Equipment Inspection And Research Institute, Beijing 100029, China
  • 4Institute of Microelectronics, Chinese Academy of Science, Beijing 100029, China
  • show less
    Figures & Tables(8)
    Experimental setup for continuous QCL gas detection system
    Experimental setup for pulsed QCL gas detection system
    Received signal waveforms of absorption peak of 4×10-5 m nitric oxide in the two systems. (a) Pulsed QCL system;(b) continuous QCL system
    Measurement curves of nitric oxide for two systems. (a) Pulsed QCL system; (b) continuous QCL system
    Measured results of background noise introduced by Etalon interference effect in two systems
    • Table 1. Analytical performance comparison of two systems

      View table

      Table 1. Analytical performance comparison of two systems

      ParameterPulsed QCLContinuous QCL
      Measurement range0‒14000‒300
      Linear range0‒5000‒300
      Limit of detection4.612.11
    • Table 2. Influence on measurement baseline by optical noise and electrical noise

      View table

      Table 2. Influence on measurement baseline by optical noise and electrical noise

      ParameterPulsed QCL systemContinuous QCL system
      Average value of etalon46.017.1
      Optical noise16.913.9
      Electrical noise1.080.70
    • Table 3. Comparison of other performance of the two systems

      View table

      Table 3. Comparison of other performance of the two systems

      ParameterPulsed QCL systemContinuous QCL system
      Minimal bandwidth of photodetectors /MHz1200.2
      Spectrum resolution /nm0.320.01
      Tuning range /nm2028
      Peak optical power /mW18060
      Time for single measurement /s0.01(1000 average)0.2
      Average power consumption /W0.312.0
    Tools

    Get Citation

    Copy Citation Text

    Zhenhan Li, Yue Yu, Rui Guo. Performance Comparison of Pulsed and Continuous Quantum Cascade Lasers in Nitric Oxide Detection[J]. Laser & Optoelectronics Progress, 2021, 58(23): 2330001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Feb. 23, 2021

    Accepted: Mar. 24, 2021

    Published Online: Nov. 19, 2021

    The Author Email: Li Zhenhan (li.zh@reyintech.com.cn)

    DOI:10.3788/LOP202158.2330001