Journal of Radiation Research and Radiation Processing, Volume. 41, Issue 4, 040303(2023)

Dose distribution alignment method based on rigid registration

Jian WEI1...2, Weicheng ZHANG1,2, Bing YAN3, Chunhua CHEN1 and Hui WANG1,* |Show fewer author(s)
Author Affiliations
  • 1Hefei Institutes of Physical Science,Chinese Academy of Sciences, Hefei 230031, China
  • 2University of Science and Technology of China, Hefei 230026, China
  • 3The First Affiliated Hospital of University of Science and Technology of China, Hefei 230001, China
  • show less
    Figures & Tables(7)
    Basic framework of image registration
    Examples of small-field radiation (a) and large-field radiation (b)
    Flow chart of data analysis
    Gamma passing rate results of different pass standards: (a1) small-field radiation measurement (3%/3 mm); (a2) large-field radiation measurement (3%/3 mm); (b1) small-field radiation measurement (3%/2 mm); (b2) large-field radiation measurement (3%/2 mm); (c1) small-field radiation measurement (2%/2 mm); (c2) large-field radiation measurement (2%/2 mm)(color online)
    • Table 1. Results of gamma (3%/3 mm) pass rate calculated directly after small-field radiation measurement, verisoft alignment and rigid registration

      View table
      View in Article

      Table 1. Results of gamma (3%/3 mm) pass rate calculated directly after small-field radiation measurement, verisoft alignment and rigid registration

      病例

      Case

      010203040506070809101112131415161718
      O / %10097.898.798.510098.61001009999.29610095.398.710098.997.8100
      V / %10097.810098.598.598.61001009899.294.110096.599.710098.998.9100
      I/ %10097.810098.510098.61001009999.294.110096.299.310098.998.9100
      OVE / %0.00.01.30.01.50.00.00.01.00.02.00.01.31.00.00.01.10.0
      OIE / %0.00.01.30.00.00.00.00.00.00.02.00.00.90.60.00.01.10.0
    • Table 2. Results of gamma (3%/3 mm) pass rate calculated directly after large-field radiation measurement, verisoft alignment and rigid registration

      View table
      View in Article

      Table 2. Results of gamma (3%/3 mm) pass rate calculated directly after large-field radiation measurement, verisoft alignment and rigid registration

      病例

      Case

      010203040506070809101112131415161718
      O / %94.19690.39897.895.795.899.795.896.310091.794.210010099.699.598.6
      V / %57.265.862.677.277.987.792.195.971.593.194.18457.288.876.97692.897.4
      I/ %95.595.591.198.497.895.796.799.496.197.410091.795.210010099.699.598.6
      OVE / %39.231.530.721.220.38.43.93.825.43.35.98.439.311.223.123.76.71.2
      OIE / %1.50.50.90.40.00.00.90.30.31.10.00.01.10.00.00.00.00.0
    • Table 3. Average error of the different gamma pass rates of the small-field radiation measurement and the large-field radiation measurement of the two alignment gamma results and the direct calculation results

      View table
      View in Article

      Table 3. Average error of the different gamma pass rates of the small-field radiation measurement and the large-field radiation measurement of the two alignment gamma results and the direct calculation results

      测量类型

      Measurement type

      对齐方式 / %

      Alignment method

      3%/3 mm3%/2 mm2%/2 mm

      小野测量

      Small-field radiation measurement

      MOVE0.50.40.8
      MOIE0.30.60.5
      大野测量MOVE17.123.328.3
      Large-field radiation measurementMOIE0.41.12.3
    Tools

    Get Citation

    Copy Citation Text

    Jian WEI, Weicheng ZHANG, Bing YAN, Chunhua CHEN, Hui WANG. Dose distribution alignment method based on rigid registration[J]. Journal of Radiation Research and Radiation Processing, 2023, 41(4): 040303

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jan. 23, 2022

    Accepted: Mar. 7, 2022

    Published Online: Sep. 21, 2023

    The Author Email:

    DOI:10.11889/j.1000-3436.2022-0010

    Topics