Journal of Applied Optics, Volume. 45, Issue 6, 1164(2024)

Field calibration method of infrared warning equipment

Dong WANG*... Hengwei ZHANG, Qianjin ZOU, Jinchao PENG and Yaxu LU |Show fewer author(s)
Author Affiliations
  • Unit 63891 of PLA, Luoyang 471003, China
  • show less
    Figures & Tables(12)
    SR800 extended surface blackbody
    Calibration results of short-range direct extended source method
    Blackbody infrared image at 37 m
    Calibration results of long-distance small source method
    Measurement error curves
    • Table 1. Parameters of SR800 extended surface blackbody

      View table
      View in Article

      Table 1. Parameters of SR800 extended surface blackbody

      参 数参数值
      温度范围/℃−40~170
      温度精度/℃±0.02
      辐射面积/mm2300×300
      发射率0.97
    • Table 2. Two points calibration results of short-range direct extended source method

      View table
      View in Article

      Table 2. Two points calibration results of short-range direct extended source method

      标定温度/℃10,2020,3030,4040,50
      响应度α232.3231.5218.4222.2
      偏移值D05873.85885.06112.56034.1
    • Table 3. Two points calibration results of long-distance small source method (37 m)

      View table
      View in Article

      Table 3. Two points calibration results of long-distance small source method (37 m)

      标定温度/℃10,2020,3030,4040,50
      响应度α193.0206.3194.4199.2
      偏移值D06100.55910.66117.76018.1
    • Table 4. Two points calibration results of long-distance small source method (63 m)

      View table
      View in Article

      Table 4. Two points calibration results of long-distance small source method (63 m)

      标定温度/℃10,2020,3030,4040,50
      响应度α179.9176.8193.9173.2
      偏移值D05727.75772.15475.35907.5
    • Table 5. Measurement errors of short-range direct extended source method

      View table
      View in Article

      Table 5. Measurement errors of short-range direct extended source method

      温度/℃黑体辐射亮度/ W∙m−2∙sr−1反演的辐射亮度/ W∙m−2∙sr−1相对 误差/%
      1011.6310.97−5.67
      2014.3013.00−9.09
      3017.3615.36−11.52
      4020.8118.08−13.12
      5024.6720.79−15.73
    • Table 6. Measurement errors of long-distance small source method (37 m)

      View table
      View in Article

      Table 6. Measurement errors of long-distance small source method (37 m)

      温度/℃黑体辐射亮度/ W∙m−2∙sr−1反演的辐射亮度/ W∙m−2∙sr−1相对 误差/%
      1011.6312.184.75
      2014.3014.602.07
      3017.3617.12−1.38
      4020.8120.19−2.98
      5024.6723.26−5.72
    • Table 7. Measurement errors of long-distance small source method (63 m)

      View table
      View in Article

      Table 7. Measurement errors of long-distance small source method (63 m)

      温度/℃黑体辐射亮度/ W∙m−2∙sr−1反演的辐射亮度/ W∙m−2∙sr−1相对 误差/%
      1011.6315.0829.66
      2014.3017.5923.01
      3017.3620.5118.15
      4020.8123.8714.70
      5024.6727.2210.34
    Tools

    Get Citation

    Copy Citation Text

    Dong WANG, Hengwei ZHANG, Qianjin ZOU, Jinchao PENG, Yaxu LU. Field calibration method of infrared warning equipment[J]. Journal of Applied Optics, 2024, 45(6): 1164

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Oct. 26, 2023

    Accepted: --

    Published Online: Jan. 14, 2025

    The Author Email: WANG Dong (wd.hardman@163.com)

    DOI:10.5768/JAO202445.0601008

    Topics