Journal of Applied Optics, Volume. 45, Issue 6, 1164(2024)
Field calibration method of infrared warning equipment
The infrared radiation calibration method was introduced, the field calibration of infrared warning equipment was carried out by short-range direct extended source method and long-distance small source method, and the verification and error analysis were carried out. The results show that the relative calibration errors of the short-range direct extended source method is between 5.67% to 15.73% at the temperature range of 10 ℃~50 ℃. The relative calibration errors of the long-distance small source method at 37 m ranging from 1.38% to 5.72%, and those at 63 m ranging from 10.34% to 29.66%, indicating that the long-distance small source method at an appropriate distance is an effective method for field calibration of infrared warning equipment operating in the mode of circular scanning.