Spectroscopy and Spectral Analysis, Volume. 42, Issue 2, 405(2022)

Effect of Wavelength Drift on PLSR Calibration Model of Near-Infrared Spectroscopy

Qi-peng LU1、*, Dong-min WANG2、2; *;, Yuan SONG1、1; *;, Hai-quan DING3、3;, and Hong-zhi GAO3、3;
Author Affiliations
  • 11. State Key Laboratory of Applied Optics Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China
  • 22. School of Food Science and Technology, Henan University of Technology, Zhengzhou 450001, China
  • 33. Guang Dong Spectrastar Instruments Co., Ltd., Guangzhou 510000, China
  • show less

    Partial least squares regression (PLSR) calibration model will be effect by the wavelength change of a single instrument at a different time and the wavelength consistency of multiple instruments. The process of near-infrared spectroscopy analysis, this problems can be unified as the effect of wavelength drift on chemometric calibration model. In this paper, taking the analysis of crude protein in wheat flour as an example, two calibration models I and II were established by partial least squares regression (PLSR) method within different spectral regions. Different types and amplitudes of wavelength shift information were generated by computer and superimposed into the spectra of the validation set to produce wavelength shift information relative to the spectra of calibration set, the effect of wavelength drift on PLSR calibration model was studied by adding different types and amplitude of wavelength drift information to the spectra of the validation set samples. The results show that the RMSEP of every model is no more than 0.3% and the corresponding Rp is no less than 0.98 when there is no wavelength drift information in the spectra of validation set samples. When the wavelength drift at different wavelengths is constant, the RMSEP increases as the wavelength drift amplitude increases, the RMSEP increases to 3.69% when the wavelength drift is -32 cm-1, and the Rp is almost constant; When the wavelength drift varies randomly at different wavelengths, the prediction results of model II based on long wavelength regions are almost not affected. The model II is corrected by a series of spectra added to the calibration set with different wavelength drift information, the RMSEP of the corrected model is 0.3%, The influence of wavelength drift information on RMSEP has been almost eliminated, but the number of regression factors used to establish corredted model increases significantly from 3 to 8, the robustness of the model varies greatly. In general, the RMSEP can be polished by correcting the prediction results to ensure the accuracy of the analysis results if the amplitude of wavelength drift is slight. This study provides an experimental basis for determining the design instrument parameters and operating procedures to improve the reliability of NIR analysis results.

    Tools

    Get Citation

    Copy Citation Text

    Qi-peng LU, Dong-min WANG, Yuan SONG, Hai-quan DING, Hong-zhi GAO. Effect of Wavelength Drift on PLSR Calibration Model of Near-Infrared Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 405

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jan. 15, 2021

    Accepted: Apr. 6, 2021

    Published Online: Apr. 2, 2022

    The Author Email: LU Qi-peng (luqipeng@126.com)

    DOI:10.3964/j.issn.1000-0593(2022)02-0405-05

    Topics