NUCLEAR TECHNIQUES, Volume. 47, Issue 2, 020401(2024)

Development of the high-resolution neutron scintillator detector for the very small-angle neutron scattering spectrometer

Huiyin LIU1,2,3, Jie YANG1、*, Chang HUANG2,3, Bin TANG2,3、**, Shihui ZHOU2,3, Xiaojie CAI2,3, Xiuku WANG2,3, Lixin ZENG2,3, Xiuping YUE2,3, Shaojia CHEN2,3, Hong XU2,3, Dawei GUO2,3, Xu CHEN2,3, and Zhijia SUN2,3
Author Affiliations
  • 1Zhengzhou University, Zhengzhou 450001, China
  • 2Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China
  • 3Spallation Neutron Source Science Center, Dongguan 523803, China
  • show less
    Figures & Tables(19)
    Schematic of VSANS spectrometer's layout
    Schematics of the detector's head (a) and mechanical structure (b)
    Movable platform for the optical test
    Plot of the optical attenuation length
    Principle diagram of WLSF's mechanical bending loss test
    Spectral comparison diagrams before and after mechanical bending (in air)
    Principle diagram of the SiPM's gain test
    Charge spectrum of the SiPM
    Plot of SiPM's gain test results
    Plots of SiPM's thermal noise test results (a) The variation of Hamamatsu's SiPM output signal with increasing temperature, (b) The variation of Sensl's SiPM output signal with increasing temperature
    Snapshot of the detector prototype
    Photo self-developed readout electronics boards
    Schematic diagram of the detection efficiency experiment
    Snapshot of boron-containing aluminum plate with "CSNS" slits
    Comparison plot of the neutron wavelength spectra
    Plot of the detection efficiency test results
    Plot of the two-dimensional imaging test result
    • Table 1. Key performance parameters of the SiPM

      View table
      View in Article

      Table 1. Key performance parameters of the SiPM

      型号

      Type

      Hamamatsu

      S13363-3050

      Sensl

      MicroFC-30035

      增益Gain3.4×1064×106
      工作偏压Over voltage2.5 V2.5 V
      击穿电压Breakdown voltage52.5 V24.5 V
      温漂系数Temperature coefficient55 mV·℃-122.3 mV·℃-1
      灵敏面积Sensitive area3 mm×3 mm3 mm×3 mm
    • Table 2. Parameters of key components

      View table
      View in Article

      Table 2. Parameters of key components

      器件名称

      Device name

      生产厂家

      Manufacturer

      型号

      Parameters

      闪烁屏 Scintillation screenScintacor6LiF/ZnS(Ag)⁓2:1/400 μm
      波移光纤WLSFKurarayY-11(200)/0.5 mm
      硅光电倍增管SiPMSenslMicroFC-30035-SMT
      电子学ElectronicsSelf-developedSelf-developed
    Tools

    Get Citation

    Copy Citation Text

    Huiyin LIU, Jie YANG, Chang HUANG, Bin TANG, Shihui ZHOU, Xiaojie CAI, Xiuku WANG, Lixin ZENG, Xiuping YUE, Shaojia CHEN, Hong XU, Dawei GUO, Xu CHEN, Zhijia SUN. Development of the high-resolution neutron scintillator detector for the very small-angle neutron scattering spectrometer[J]. NUCLEAR TECHNIQUES, 2024, 47(2): 020401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jun. 26, 2023

    Accepted: --

    Published Online: Apr. 24, 2024

    The Author Email: YANG Jie (杨洁), TANG Bin (唐彬)

    DOI:10.11889/j.0253-3219.2024.hjs.47.020401

    Topics