Spectroscopy and Spectral Analysis, Volume. 42, Issue 6, 1666(2022)

Fourier Transform Near-Infrared Spectral System Based on Laser-Driven Plasma Light Source

Figures & Tables(12)
Spectral intensity and coverage band comparison of different sources
Photosignal and frequency of LDLS
Optical path of near-infrared Fourier transform spectrometer
Near-infrared Fourier transform spectrometer system(a): Light source high-frequency modulated by using chopper; (b): LDLS system
Detector background noise without light signal input
The spectral comparison results with different integration times of the lock-in amplifier(a): Integration time=20 ms; (b): Integration time=10 ms; (c): Integration time=5 ms;(d): Integration time=1 ms; (e): Integration time=0.5 ms; (f): SNRI vs integration time
Fourier transform interference spectral results with LDLS(a): Without demodulation; (b): With demodulation
Spectral SNRI before and after replacing the light source
The near-infrared standard sample absorption spectra measured with LDLS Fourier transform spectral system(a): Fourier transformation spectra with and without standard sample; (b): near-infrared Fourier transformation spectra with background removed
  • Table 1. The equipment and model required for the experiment

    View table
    View in Article

    Table 1. The equipment and model required for the experiment

    NameMain ParametersProduct ModelManufacturers
    LDLSRadiance>40 mW·mm-2·sr·nm
    λ=170~2 100 nm
    EQ77Energetiq
    He-Ne LaserP=21 mW
    λ=632.8 nm
    22037146/1145PEdmund
    Lock-in amplifierSample Rate210 MSa·s-1
    The frequency of work 0.7 μHz~50 MHz
    HF2LIZurich Instruments
    Halogen tungsten light sourceλ=360~2600 nm
    P>10 mW
    SLS201L/MThorlabs
    ChopperChopper frequency 10 kHzMC1F10HPThorlabs
    InGaAs Photodectorλ=900~2 570 nmPDA10DT-ECThorlabs
    Position sensing detectorλ=400~1 050 nmPDQ30CThorlabs
    Off-axis mirrorf=100 mmMPD249-M01Thorlabs
    Data acquisition cardSample Rate 3.57 MSa·s-1PCIe-6374NI
  • Table 2. Near-infrared peak and bandwidth of standard sample

    View table
    View in Article

    Table 2. Near-infrared peak and bandwidth of standard sample

    Serial
    number
    Peak
    /nm
    Bandwidth
    /nm
    Peak
    /nm
    Bandwidth
    /nm
    11 070.221 068.35
    21 220.321 220.55
    31 363.721 362.95
    41 464.721 464.65
    51 524.021 524.05
    61 938.421 939.05
  • Table 3. Comparison of standard peaks and test peaks

    View table
    View in Article

    Table 3. Comparison of standard peaks and test peaks

    Standard
    peaks 1/nm
    Standard
    peaks 2/nm
    Test peaks
    /nm
    1 070.21 068.31 068.5
    1 220.31 220.51 220.4
    1 363.71 362.91 363.3
    1 464.71 464.61 464.8
    1 524.01 524.01 523.9
    1 938.41 939.01 938.7
Tools

Get Citation

Copy Citation Text

. Fourier Transform Near-Infrared Spectral System Based on Laser-Driven Plasma Light Source[J]. Spectroscopy and Spectral Analysis, 2022, 42(6): 1666