Chinese Optics Letters, Volume. 22, Issue 3, 031701(2024)

Stimulated emission–depletion-based point-scanning structured illumination microscopy

Lei Wang1,2, Meiting Wang1, Luwei Wang3, Xiaomin Zheng3, Jiajie Chen3、*, Wenshuai Wu3, Wei Yan3, Bin Yu3, Junle Qu3, Bruce Zhi Gao4, and Yonghong Shao3、**
Author Affiliations
  • 1Guangdong Key Laboratory for Biomedical Measurements and Ultrasound Imaging, National-Regional Key Technology Engineering Laboratory for Medical Ultrasound, School of Biomedical Engineering, Shenzhen University Medical School, Shenzhen 518060, China
  • 2Key Laboratory of Opto-electronic Information Science and Technology of Jiangxi Province, Nanchang Hangkong University, Nanchang 330063, China
  • 3College of Physics and Optoelectronics Engineering, Key Laboratory of Optoelectronic Devices and Systems of Ministry of Education and Guangdong Province, Shenzhen University, Shenzhen 518060, China
  • 4Department of Bioengineering and COMSET, Clemson University, Clemson SC 29634, US
  • show less
    Figures & Tables(3)
    (a) Schematic diagram of STED-psSIM; (b) relationship among the excitation beam, the depletion beam, and the fluorescence signal in the temporal and spatial domain.
    Comparison of imaging resolutions of confocal microscopy, STED microscopy, and STED-psSIM. (a) Images of 23 nm fluorescent beads captured by the confocal microscopy (lower left), STED microscopy (middle), and STED-psSIM (upper right) separated by a dashed green line; (b)–(d) magnified view of the area encircled by the green dashed box in (a): (b) confocal microscopy, (c) STED microscopy, and (d) STED-psSIM[28]; (e)–(g) observable spatial frequency regions obtained by Fourier transforming images in (b)–(d); (h), (i) normalized intensity profiles of fluorescence beads along the (h) blue and (i) yellow dashed lines in (b)–(d), respectively. The profiles show average FWHMs of 193, 117, and 74 nm. Scale bars, 1 µm in (a) and 0.2 µm in (b)–(d).
    Resolution improvement in STED-psSIM for biological imaging. (a) Images of NPC and (b) labeled actin in huFIB cells captured by confocal microscope (lower left), STED (middle), and STED-SIM (upper right) separated by two dashed green lines; (c) and (e) magnified view of the area encircled by the dashed box in (a) and (b); (d) and (f) normalized intensity profiles along the green dashed lines in (c) and (e), respectively; (g) SBR of the two biological samples in the three methods. Scale bars, 1 µm in (a), (b), 0.4 µm in (c), and 0.2 µm in (e).
    Tools

    Get Citation

    Copy Citation Text

    Lei Wang, Meiting Wang, Luwei Wang, Xiaomin Zheng, Jiajie Chen, Wenshuai Wu, Wei Yan, Bin Yu, Junle Qu, Bruce Zhi Gao, Yonghong Shao. Stimulated emission–depletion-based point-scanning structured illumination microscopy[J]. Chinese Optics Letters, 2024, 22(3): 031701

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Biophotonics

    Received: Oct. 6, 2023

    Accepted: Oct. 26, 2023

    Published Online: Mar. 14, 2024

    The Author Email: Jiajie Chen (cjj@szu.edu.cn), Yonghong Shao (shaoyh@szu.edu.cn)

    DOI:10.3788/COL202422.031701

    Topics