Chinese Journal of Quantum Electronics, Volume. 41, Issue 3, 496(2024)

Detection method of rare earth elements based on LIBS technology and multivariate linear calibration

XU Zhishuai1...2,*, XIAO Yihao1, LIU Li1, and HAO Zhongqi12 |Show fewer author(s)
Author Affiliations
  • 1Key Laboratory for Optoelectronic Information Perception and Instrumentation of Jiangxi Province, Nanchang Hangkong University, Nanchang 330063, China
  • 2Key Laboratory of Non-damage Inspection and Measurement Technology of Ministry of Education, Nanchang Hangkong University, Nanchang 330063, China
  • show less
    Figures & Tables(10)
    Schematic diagram of the LIBS experimental device
    Variation of spectral line intensity (a), SNR (b) with defocus distance
    Variation of spectral line intensity (a) and SNR (b) with acquisition delay time
    Variation of spectral line intensity (a), SNR (b) with pulse energy
    Selected characteristic spectral lines of La (a), Gd (b), Y (c), Yb (d)
    Relationship between La predicted and reference values of LIBS
    Correlation coefficients of multiple linear regression equations
    Variation of RMSE of multiple linear regression models of four rare earth elements
    • Table 1. Sample mixing ratio and element concentration

      View table
      View in Article

      Table 1. Sample mixing ratio and element concentration

      Sample No.Raw material quality/mgElement concentration/%
      La2O3Gd2O3Y2O3Yb2O3CaCO3LaGdYYb
      1106.6740.0093.3310.003083.332.731.042.200.26
      272.5037.5062.5015.002312.502.471.301.970.53
      343.3330.0036.6715.001541.672.221.561.730.79
      428.7526.2523.7515.001156.251.961.821.501.05
      520.0024.0016.0015.00925.001.712.081.261.32
      619.6231.1515.0020.771067.311.452.341.021.58
      721.0045.0015.0031.501387.501.192.600.791.84
      823.5775.0015.0051.431977.860.943.040.552.11
      930.00142.5015.00101.253461.250.683.300.312.37
      1025.00165.0010.00150.004650.000.432.860.162.63
      1110.00145.0025.00175.004645.000.172.520.393.07
      1215.0097.5030.00142.503465.000.342.260.633.34
      1315.0049.2923.5770.711984.290.602.000.872.90
      1415.0030.0021.0037.501396.500.851.741.102.20
      1515.0019.6219.6225.381074.231.111.471.341.93
      1617.1415.0021.4318.21999.641.361.211.571.49
      1725.9115.0031.3617.731273.641.620.951.811.14
      1841.2515.0048.7515.001755.001.880.692.050.70
      1975.0015.0090.0015.002805.002.130.432.360.44
      20140.0010.00175.0010.004665.002.390.172.760.18
    • Table 2. Selection of spectral lines for multiple linear regression

      View table
      View in Article

      Table 2. Selection of spectral lines for multiple linear regression

      ModelSelected La spectral line/nmSelected Gd spectral line/nmSelected Y spectral line/nmSelected Yb spectral line/nm
      Unary394.91342.25437.49369.42
      Binary

      394.91

      433.37

      342.25

      376.84

      437.49

      324.23

      369.42

      328.94

      Ternary

      394.91

      433.37

      412.32

      342.25

      376.84

      335.05

      437.49

      324.23

      371.03

      369.42

      328.94

      289.14

    Tools

    Get Citation

    Copy Citation Text

    Zhishuai XU, Yihao XIAO, Li LIU, Zhongqi HAO. Detection method of rare earth elements based on LIBS technology and multivariate linear calibration[J]. Chinese Journal of Quantum Electronics, 2024, 41(3): 496

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Special Issue on Key Technologies and Applications of LIBS

    Received: Jan. 29, 2024

    Accepted: --

    Published Online: Jul. 17, 2024

    The Author Email: XU Zhishuai (2367985761@qq.com)

    DOI:10.3969/j.issn.1007-5461.2024.03.011

    Topics