Journal of Applied Optics, Volume. 44, Issue 6, 1355(2023)

Development of transmission visibility meter calibration device and analysis of measurement results

Qing LI1... Yingxin ZHANG1, Fan ZHANG2, Hongda TAI3, Huayang HE4 and Yi LI1,* |Show fewer author(s)
Author Affiliations
  • 1Key Laboratory of Metrological Optics and Application for State Market Regulation, Shaanxi Institute of Metrology Science, Xi'an 710100, China
  • 2Institute of Length Metrology, Jiangsu Institute of Metrology, Nanjing 210023, China
  • 3Department of Aviation Meteorology, Civil Aviation University of China, Tianjin 300300, China
  • 4Research Institute of Highway Ministry of Transport, Beijing 100088, China
  • show less
    Figures & Tables(10)
    Schematic diagram of principle of transmissometer
    Schematic diagram of transmissometer calibration device
    Analysis of filter front-back linearity
    Linearity fitting curves of mental filter
    Linearity fitting curves of glass filter
    Influence of circle center offset on visibility measurement
    • Table 1. Influence of filter transmittance a and central angle ratio p on visibility VR

      View table
      View in Article

      Table 1. Influence of filter transmittance a and central angle ratio p on visibility VR

      apTUrel(k=2)VR/m
      注:基线长度L为1 m。
      0.2501/100.925 00.000 650
      0.7501/100.975 00.000 6155
      0.9501/100.995 00.000 6780
      0.9501/50.990 00.001 2389
      0.9501/200.997 50.000 31 563
    • Table 2. Theoretical values of visibility corresponding to two sets of filters

      View table
      View in Article

      Table 2. Theoretical values of visibility corresponding to two sets of filters

      序号圆心角/(°)透射比能见度理论值/m
      L=1L=2
      A2.80.984 4249.5499.1
      B1.80.990 0389.2778.5
      C1.70.999 24 6009 201
      D2.00.999 03 9107 820
      E3.10.998 52 5225 044
      F2.20.998 93 5547 109
      G3.50.998 32 2334 467
    • Table 3. Filter actual center and relative error of visibility

      View table
      View in Article

      Table 3. Filter actual center and relative error of visibility

      序号圆心角/(°)圆心角误差/(°)透射比能见度/m相对误差/%
      L=1L=2
      A2.820.020.984 3247.7495.4−0.71
      B1.830.030.989 8382.8765.6−1.65
      C1.67−0.030.999 24 6839 3661.80
      D1.99−0.010.999 03 9297 8590.50
      E3.07−0.030.998 52 5465 0930.98
      F2.19−0.010.998 93 5717 1410.46
      G3.47−0.030.998 32 2524 5060.87
    • Table 4. Sources of uncertainty and evaluation results

      View table
      View in Article

      Table 4. Sources of uncertainty and evaluation results

      不确定度来源不确定度/%类别
      标准器测量重复性引入的不确定度分量0.22A
      切割光线角度误差引入的不确定度分量1.06B
      上级量传引入的不确定度2.05B
      滤光片透射比引入的不确定度分量2.50B
      实验室内大气扰动、杂散光干扰和背景辐射等引入的不确定度1.00B
      相对合成标准不确定度3.55
      相对扩展不确定度(k=27.10
    Tools

    Get Citation

    Copy Citation Text

    Qing LI, Yingxin ZHANG, Fan ZHANG, Hongda TAI, Huayang HE, Yi LI. Development of transmission visibility meter calibration device and analysis of measurement results[J]. Journal of Applied Optics, 2023, 44(6): 1355

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: May. 31, 2023

    Accepted: --

    Published Online: Mar. 12, 2024

    The Author Email: LI Yi (李奕)

    DOI:10.5768/JAO202344.0605001

    Topics