High Power Laser and Particle Beams, Volume. 36, Issue 5, 055002(2024)

A design of B-dot calibration simulator for azimuthal transmission line

Xinjun Zhang... Weixi Luo, Yixiang Hu, Jiahui Yin and Wenyuan Zhou |Show fewer author(s)
Author Affiliations
  • State Key Laboratory of Intense Pulsed Radiation Simulation and Effect, Northwest Institute of Nuclear Technology, Xi’an 710024, China
  • show less
    Figures & Tables(10)
    Structure of azimuthal transmission line
    Structure of the off-line calibration platform
    Structural dimension machining error
    Installation eccentricity error
    Structure of B-dot
    Output deviation of 5% at different initial position
    Schematic diagram of the B-dot installation angle
    Angle deviation for output deviation of 5% at different initial angle
    • Table 1. Peak deviation of calibrated signal on the 1∶1 device

      View table
      View in Article

      Table 1. Peak deviation of calibrated signal on the 1∶1 device

      f/MHz($2I_{A_{1}} $/$I_{A_{\mathrm{in}}} $−1)/%f/MHz($2I_{A_{1}} $/$I_{A_{\mathrm{in}}} $−1)/%
      10.64623
      23.1722
      33.2824
      48.4935
      5171042
    • Table 2. Peak deviation of calibrated signal on the calibration simulator

      View table
      View in Article

      Table 2. Peak deviation of calibrated signal on the calibration simulator

      f/MHz(Ic/Iin−1)/%f/MHz(Ic/Iin−1)/%
      10.0361.2
      20.1471.6
      30.3281.8
      40.5792.8
      50.87103.8
    Tools

    Get Citation

    Copy Citation Text

    Xinjun Zhang, Weixi Luo, Yixiang Hu, Jiahui Yin, Wenyuan Zhou. A design of B-dot calibration simulator for azimuthal transmission line[J]. High Power Laser and Particle Beams, 2024, 36(5): 055002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Oct. 6, 2023

    Accepted: Mar. 15, 2024

    Published Online: May. 16, 2024

    The Author Email:

    DOI:10.11884/HPLPB202436.230341

    Topics