Opto-Electronic Engineering, Volume. 43, Issue 1, 55(2016)

All-reflective Broadband Spectroscopic Imaging Ellipsometer

JIANG Chunguang1...2,*, CHEN Yaqin2, LIU Tao2, XIONG Wei2, LI Guoguang2 and JI Feng1 |Show fewer author(s)
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    We developed a broadband spectroscopic imaging ellipsometer, which is free of chromatic aberration, by using an all-reflective focusing optical structure with special polarization control. A calibration method by measuring multiple standard samples was employed in the system calibration procedure. By applying the obtained system calibration parameters, we can determine the test sample’s spatial distributions of ellipsometric angles ψ and σ, and the film thickness after the imaging ellipsometric analysis. To test the accuracy of our home-made imaging ellipsometer, we have measured the SiO2/Si samples with the thicknesses of 3 nm~300 nm at multiple wavelengths between 200 nm and 1 000 nm. The experimental result shows that the SiO2 film thickness can be determined within the maximum relative measurement error of 6%.

    Tools

    Get Citation

    Copy Citation Text

    JIANG Chunguang, CHEN Yaqin, LIU Tao, XIONG Wei, LI Guoguang, JI Feng. All-reflective Broadband Spectroscopic Imaging Ellipsometer[J]. Opto-Electronic Engineering, 2016, 43(1): 55

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jan. 28, 2015

    Accepted: --

    Published Online: Mar. 22, 2016

    The Author Email: Chunguang JIANG (chunguangj@163.com)

    DOI:10.3969/j.issn.1003-501x.2016.01.010

    Topics