Journal of Radiation Research and Radiation Processing, Volume. 40, Issue 5, 050302(2022)

Setup error and planning target volume boundary in foam-immobilized radiotherapy for breast cancer

Yuan CHENG1... Zhenghuan LI3, Youlin CHEN2 and Suyu LI2,* |Show fewer author(s)
Author Affiliations
  • 1Fuzhou No.1 Hospital Affiliated with Fujian Medical University, Fuzhou 350000, China
  • 2Fujian Maternity and Child Health Hospital, Fuzhou 350005, China
  • 3The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou 510630, China
  • show less
    Figures & Tables(8)
    Breast cancer foam immobilize position
    CT images of Breast cancer
    Osseous registration results of mmobilized CT images and CBCT images
    Frequency distribution histogram of set-up error in X/Y/Z direction
    • Table 1. Individual set-up errorin 31 patients with breast cancer

      View table
      View in Article

      Table 1. Individual set-up errorin 31 patients with breast cancer

      患者编号

      No.

      X方向(左右)

      X direction (left-right)

      Y方向(头脚)

      Y direction (the head and foot)

      Z方向(腹背)

      Z direction (the abdomen and back)

      系统误差

      Systematic error

      随机误差

      Random error

      系统误差

      Systematic error

      随机误差

      Random error

      系统误差

      Systematic error

      随机误差

      Random error

      1-0.371.81-0.371.480.903.06
      2-3.180.641.031.64-0.280.59
      3-1.403.14-2.651.07-2.303.55
      4-0.981.400.801.174.151.39
      51.500.880.530.981.630.56
      6-1.780.850.731.350.951.04
      71.280.79-1.431.58-0.450.40
      81.800.442.071.40-0.532.25
      90.431.75-2.212.03-2.674.65
      101.300.610.100.46-2.330.59
      111.300.892.450.441.601.03
      12-3.242.22-1.722.101.581.20
      13-1.521.17-3.601.13-4.560.80
      14-1.131.03-0.101.31-0.431.90
      150.631.023.301.28-0.201.07
      160.530.680.581.761.230.85
      17-1.300.610.582.971.581.00
      180.132.682.031.12-2.134.67
      19-1.331.483.080.46-2.801.41
      200.172.00-0.300.35-2.400.44
      21-2.081.33-1.581.710.434.85
      220.001.350.651.110.151.06
      230.530.290.131.50-0.900.26
      240.681.42-0.340.96-1.241.34
      251.281.403.730.59-0.931.68
      26-1.581.130.420.800.021.37
      272.051.312.231.140.381.20
      280.332.002.930.670.601.99
      29-0.331.850.831.160.631.42
      30-0.380.83-0.171.55-0.301.53
      311.030.32-3.330.74-3.100.92
    • Table 2. Group set-up error in 31 patients with breast cancer

      View table
      View in Article

      Table 2. Group set-up error in 31 patients with breast cancer

      坐标方向

      Coordinate direction

      绝对最大值

      Absolute maximum

      绝对最小值

      Absolute minimum

      群体系统误差

      Population systematic error

      群体随机误差

      Population random error

      XX axis4.901.400.67
      YY axis6.401.890.56
      ZZ axis8.701.681.16
    • Table 3. Incidence of set-up errors in 31 breast cancer patients

      View table
      View in Article

      Table 3. Incidence of set-up errors in 31 breast cancer patients

      误差范围

      Error range

      X方向(左右)

      X direction (left-right)

      Y方向(头脚)

      Y direction (the head and foot)

      Z方向(腹背)

      Zdirection (the abdomen and back)

      ≤3 mm116(88.55)104(79.39)99(75.57)
      3~5 mm15(11.45)25(19.08)24(18.32)
      >5 mm0(0.00)2(1.53)8(6.11)
    • Table 4. Set-up errors of breast cancer patients and expand margins of PTV

      View table
      View in Article

      Table 4. Set-up errors of breast cancer patients and expand margins of PTV

      X方向(左右)

      X direction (left-right)

      Y方向(头脚)

      Y direction (the head and foot)

      Z方向(腹背)

      Z direction (the abdomen and back)

      1.401.891.68
      δ0.670.561.16
      MPTV / mm3.985.115.02
    Tools

    Get Citation

    Copy Citation Text

    Yuan CHENG, Zhenghuan LI, Youlin CHEN, Suyu LI. Setup error and planning target volume boundary in foam-immobilized radiotherapy for breast cancer[J]. Journal of Radiation Research and Radiation Processing, 2022, 40(5): 050302

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Mar. 3, 2022

    Accepted: Jun. 21, 2022

    Published Online: Nov. 3, 2022

    The Author Email: LI Suyu (lisuyu2008@163.com)

    DOI:10.11889/j.1000-3436.2022-0027

    Topics