Spectroscopy and Spectral Analysis, Volume. 40, Issue 11, 3562(2020)

Study on Cefradine Granules Component Analysis and Calibration Transfer Method Based on Near-Infrared Spectroscopy

Zi-kun ZHOU1,1、*, Chen-xi LI1,1, Zhe WANG1,1, Rong LIU1,1, Wen-liang CHEN1,1, and Ke-xin XU1,1
Author Affiliations
  • 1[in Chinese]
  • 11. State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China
  • show less
    Figures & Tables(5)
    Relation between evolution step and mean square error
    The relation between the normalized parameters (RMSEP and PR) and the number of conversion sets(a): Cefradine; (b): Cephalexin; (c): Moisture content
    • Table 1. Modeling parameters and prediction results of two spectrometers

      View table
      View in Article

      Table 1. Modeling parameters and prediction results of two spectrometers

      成分头孢拉定头孢氨苄水分
      仪器FrontierAntarisFrontierAntarisFrontierAntaris
      RMSEC/(mg·g-1)1.832.090.0890.1880.1210.147
      RMSEP/(mg·g-1)2.923.630.3040.3160.1470.189
      R20.9880.9820.6230.6110.8740.767
    • Table 2. Calibration transfer parameters and spectral differences

      View table
      View in Article

      Table 2. Calibration transfer parameters and spectral differences

      成分头孢
      拉定
      头孢
      氨苄
      水分
      窗口宽度9119
      传递后ARMS0.0750.0760.083
      Pr/%91.194.794.2
      主仪器RMSEP/(mg·g-1)2.2640.2100.131
      从仪器传递前RMSEP/(mg·g-1)8.820.4810.362
      从仪器传递后RMSEP/(mg·g-1)3.500.3140.172
      从仪器传递前RPE/%9.6752.1419.25
      从仪器传递后RPE/%4.3731.1211.67
    • Table 3. Calibration transfer and evaluation parameters of two algorithms

      View table
      View in Article

      Table 3. Calibration transfer and evaluation parameters of two algorithms

      成分头孢拉定头孢氨苄水分
      方法MCK-SMCK-SMCK-S
      ARMSun0.2510.2440.3320.3170.3450.328
      ARMS0.0750.0810.0760.0830.0830.091
      Pr/%91.189.094.793.194.292.3
      RMSEP/(mg·g-1)3.503.590.3140.2880.1720.179
    Tools

    Get Citation

    Copy Citation Text

    Zi-kun ZHOU, Chen-xi LI, Zhe WANG, Rong LIU, Wen-liang CHEN, Ke-xin XU. Study on Cefradine Granules Component Analysis and Calibration Transfer Method Based on Near-Infrared Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2020, 40(11): 3562

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Sep. 19, 2019

    Accepted: --

    Published Online: Jun. 18, 2021

    The Author Email: ZHOU Zi-kun (137779278@qq.com)

    DOI:10.3964/j.issn.1000-0593(2020)11-3562-05

    Topics