Optics and Precision Engineering, Volume. 32, Issue 18, 2744(2024)

Design of NIR-II imaging system with wide spectrum

Yang JIANG1,*... Dongxu ZHAO2, Xiyao SONG1,3, Yiqi GU4, Yan XING1 and Bo ZHENG5 |Show fewer author(s)
Author Affiliations
  • 1Changchun Institute of Optics,Fine Mechanics and Physics,Chinese Academy of Sciences , Changchun30033,China
  • 2Guangdong-Hong Kong-Macao Joint Laboratory for Intelligent Micro-Nano Optoelectronic College of Physics and Optoelectronic Engineering, Foshan University, Foshan58000,China
  • 3University of Chinese Academy of Sciences,Beijing100049,China
  • 4China-Japan Union Hospital of Jilin University, Changchun130033,China
  • 5Guangdong Polytechnic of Science and Technology, Zhuhai19090,China
  • show less
    Figures & Tables(12)
    Schematic layout of perfect lens
    Initial configuration with real lens (a) &Transverse ray fan plot (b)
    Initial configuration with cemented doublet (a) &Transverse ray fan plot (b)
    Layout of optical structure (a)& Layout ray fan plot (b)
    MTF of optical system at various target distances
    Spot Diagram of the optical system
    Field curvature &F-tan(theta) distortion
    Polychromatic diffraction through focus MTF
    Experimental system for NIR-II imaging
    • Table 1. Specifications of optical system

      View table
      View in Article

      Table 1. Specifications of optical system

      IndexParameters
      Working spectrum/μm0.9~2.3
      Resolution640×512
      Detector pixels/μm15
      F number2
      Focal length/mm40
      Focus range/mm100~+∞
      BFL/mm>50
    • Table 2. Tolerance distribution parameters for optical systems

      View table
      View in Article

      Table 2. Tolerance distribution parameters for optical systems

      TorlerancesLimits
      Radius(Fringes)3
      Thickness/mm0.02
      Surface decenter XY/mm0.01
      Surface tilt XY/(°)0.01
      Surface irregularity0.3
      Element decenter XY/mm0.02
      Element tilt XY/(°)0.1
    • Table 3. Result of Monte Carlo tolerance analysis

      View table
      View in Article

      Table 3. Result of Monte Carlo tolerance analysis

      概率MTF值
      98%0.299 380 71
      90%0.360 931 96
      80%0.401 494 56
      50%0.480 058 43
      20%0.553 043 06
      10%0.586 725 46
      2%0.632 721 63
    Tools

    Get Citation

    Copy Citation Text

    Yang JIANG, Dongxu ZHAO, Xiyao SONG, Yiqi GU, Yan XING, Bo ZHENG. Design of NIR-II imaging system with wide spectrum[J]. Optics and Precision Engineering, 2024, 32(18): 2744

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Apr. 28, 2024

    Accepted: --

    Published Online: Nov. 18, 2024

    The Author Email: JIANG Yang (le_zhi@yeah.net)

    DOI:10.37188/OPE.20243218.2744

    Topics