Non-linear Calibration Temperature Point Selection Method for Infrared Spectral Imager

Yanheng WANG1,2, Yun LI1, Can YU1,2, Xiangyu GAO1,2, and Shuang WANG1,2、*
Author Affiliations
  • 1Key Laboratory of Spectral Imaging Technology,Xi'an Institute of Optics and Precision Mechanics,Chinese Academy of Sciences,Xi'an 710119,China
  • 2University of Chinese Academy of Sciences,Beijing 100049,China
  • show less
    Figures & Tables(13)
    Imaging process
    Calibration of various spectral bands at the same temperature point
    Simulation of nonlinear detector response relationships
    Simulation of the s-e relationship in the detector response to 17 bands
    Comparison of division results(relative standard deviation 1%)
    Comparison of the results of the two methods divided into 10 segments
    Nonlinear response curves for 17 spectral bands classified by the standard deviation method
    Comparison of calibration results of different number of calibration temperature points
    Experiment for acquiring infrared radiation
    Standard deviation method division and linear fitting result
    • Table 1. Comparison of the results from both segmentation methods with identical segmentation requirements

      View table
      View in Article

      Table 1. Comparison of the results from both segmentation methods with identical segmentation requirements

      MethodsNumber of segmentsNumber of calibration pointsNon-linearity/%
      Dichotomy1122<1%
      Standard deviation method714<1%
    • Table 2. Comparison of the results obtained from the two methods divided into 10 segments

      View table
      View in Article

      Table 2. Comparison of the results obtained from the two methods divided into 10 segments

      Non-linearity/%Uniform divisionStandard deviation method
      Maximum1.060.40
      Averages0.440.35
    • Table 3. Comparison of non-linearity calculation results for calibrating different numbers of temperature points

      View table
      View in Article

      Table 3. Comparison of non-linearity calculation results for calibrating different numbers of temperature points

      Non-linearity/%90 calibration temperature points18 calibration temperature points10 calibration temperature points
      Maximum0.912.463.34
      Averages0.550.730.98
    Tools

    Get Citation

    Copy Citation Text

    Yanheng WANG, Yun LI, Can YU, Xiangyu GAO, Shuang WANG. Non-linear Calibration Temperature Point Selection Method for Infrared Spectral Imager[J]. Acta Photonica Sinica, 2024, 53(11): 1130003