Acta Photonica Sinica, Volume. 53, Issue 2, 0212004(2024)

Method of Calibrating Target Distance Using Dual Zone-block Light Screen Velocity Measuring Device

Jing LI1, Zhonghui SUN2, Jinping NI1,3、*, and Hui TIAN1
Author Affiliations
  • 1School of Opto-electronic Engineering,Xi'an Technological University,Xi'an 710021,China
  • 2Chinese People's Liberation Army 63853 Unit,Baicheng 137001,China
  • 3Shaanxi Province Key Laboratory of Photoelectric Measurement and Instrument Technology,Xi'an 710021,China
  • show less
    Figures & Tables(13)
    Schematic of velocity measurement by single zone-block device
    Schematic of calibration by single zone-block device
    Schematic of velocity measurement by dual zone-block device
    Schematic of calibration by dual zone-block device
    Curve of ΔSGH as a function of d
    Curve of ΔSGH as a function of L
    Schematic of test layout
    Test site of calibration with dual zone-block device
    • Table 1. Results of calibration by single zone-block device

      View table
      View in Article

      Table 1. Results of calibration by single zone-block device

      Firing sequencev23/(m·s-1tGH/μstHK/μsSGH1/mmSHK1/mm
      Average value74.8874.87
      1117.13638.6638.674.8074.80
      2154.60485.6483.675.0774.76
      3164.91454.0454.074.8774.87
      4169.84440.8441.674.8675.00
      5166.29449.8450.674.8074.93
    • Table 2. Results of velocity measurement used single zone-block calibration method

      View table
      View in Article

      Table 2. Results of velocity measurement used single zone-block calibration method

      Firing sequencev23/(m·s-1vGK1/(m·s-1ΔvGK1/(m·s-1δvGK1/(‰)
      Average value0.221.55
      1117.13117.020.110.95
      2154.60154.240.362.31
      3164.91164.690.221.31
      4169.84169.580.251.50
      5166.29166.010.281.69
    • Table 3. Times value sequence of projectile pass through screen

      View table
      View in Article

      Table 3. Times value sequence of projectile pass through screen

      Firing sequencet12/μst34/μst1G/μst1H/μstGH'/μstHK'/μs
      1862 2881 6153 23158 52528.6528.6
      2127 79130 44227 05234 88782.4782.6
      3883 9903 5157 06162 49541.6541.8
      4100 14102 28177 95184 09613.4613.6
      5808 9826 9143 76148 71495.8496.0
    • Table 4. Results of calibration by dual zone-block device

      View table
      View in Article

      Table 4. Results of calibration by dual zone-block device

      Firing sequenceSGH2/mmSHK2/mmΔS1/mmΔS2/mm
      Average value74.9874.980.060.06
      175.0074.960.060.06
      274.9775.010.050.05
      374.9774.960.060.06
      474.9874.970.060.06
      574.9974.990.060.06
    • Table 5. Results of velocity measurement used dual zone-block calibration method

      View table
      View in Article

      Table 5. Results of velocity measurement used dual zone-block calibration method

      Firing sequencev23/(m·s-1vGK2/(m·s-1ΔvGK2/(m·s-1δvGK2/(‰)
      Average value0.050.37
      1117.13117.210.080.66
      2154.60154.490.110.70
      3164.91164.960.050.30
      4169.84169.860.020.12
      5166.29166.280.010.07
    Tools

    Get Citation

    Copy Citation Text

    Jing LI, Zhonghui SUN, Jinping NI, Hui TIAN. Method of Calibrating Target Distance Using Dual Zone-block Light Screen Velocity Measuring Device[J]. Acta Photonica Sinica, 2024, 53(2): 0212004

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jul. 20, 2023

    Accepted: Oct. 23, 2023

    Published Online: Mar. 28, 2024

    The Author Email: NI Jinping (nijp6504@263.com)

    DOI:10.3788/gzxb20245302.0212004

    Topics