Optics and Precision Engineering, Volume. 32, Issue 14, 2189(2024)
Design of automatic image measuring system Research on dynamic and high-sensitive detection of SF6 discharge decomposition products by fluorescence-suppressed fiber-enhanced Raman spectroscopy
Fig. 1. Schematic diagram of the fluorescence-suppressed FERS detection system of SF6 discharge decomposition products
Fig. 3. Raman spectra of air acquired by FERS gas detection system under different filtering methods
Fig. 6. Concentration calibration of 7 kinds of SF6 discharge decomposition products
Fig. 7. Raman spectra of SF6 discharge decomposition products measured continuously in 168 h
Fig. 8. Raman spectrum of nearby region of HF characteristic Raman peak after 168 h of experiment
Fig. 9. Concentration changes of SF6 discharge decomposition products in 168 h
|
|
|
|
Get Citation
Copy Citation Text
Tongqin RAN, Fu WAN, Quan ZHOU, Weiping KONG, Qiang LIU, Yingkai LONG, Weigen CHEN. Design of automatic image measuring system Research on dynamic and high-sensitive detection of SF6 discharge decomposition products by fluorescence-suppressed fiber-enhanced Raman spectroscopy[J]. Optics and Precision Engineering, 2024, 32(14): 2189
Category:
Received: Apr. 15, 2024
Accepted: --
Published Online: Sep. 27, 2024
The Author Email: WAN Fu (wanfuhappy@163.com)