Infrared and Laser Engineering, Volume. 50, Issue 6, 20200433(2021)

Design of PGP imaging spectrometer with eliminating spectral line bending

Zhikun Wu1,2,3,4, Entao Shi1,2,3、*, and Yongmei Wang1,2,3,4
Author Affiliations
  • 1National Space Science Center, Chinese Academy of Sciences, Beijing 100190, China
  • 2Beijing Key Laboratory of Space Environment Exploration, Beijing 100190, China
  • 3Key Laboratory of Science and Technology on Space Environmental Situation Awareness, Chinese Academy of Sciences, Beijing 100190, China
  • 4University of Chinese Academy of Sciences, Beijing 100049, China
  • show less
    Figures & Tables(11)
    The geometry of incident ray
    Principle of the PGP component
    Smile distortion with x-field
    Optical layout of improved PGP imaging spectrometer system
    RMS spot with wavelength
    Diffractive MTF curves of central wavelength
    Smile distortion with normalized x-field
    Keystone distortion with wavelength
    • Table 1. Design parameters of PGP imaging spectrometer

      View table
      View in Article

      Table 1. Design parameters of PGP imaging spectrometer

      ParameterValue
      Spectral range/nm400-1000
      Spectral resolution/nm2.8
      Object spaceNA0.18
      Slit length/mm≥9
      Image width (spectral)/mm≤6.2
      Grating frequency/lp·mm−1360
    • Table 2. Parameters of the improved PGP component

      View table
      View in Article

      Table 2. Parameters of the improved PGP component

      ParameterValue
      Bragg wavelength/nm650
      Central wavelength/nm700
      Vertex angle of the first prism/(°)1.91
      Vertex angle of the second prism/(°)14.95
      Prism materialH-K9L
      Grating materialB270
      Collimating lens focal length/mm26
      Off-axis slit position/mm2.6
    • Table 3. Allowable tolerance of optical system

      View table
      View in Article

      Table 3. Allowable tolerance of optical system

      ParameterValue
      Index0.0003
      Abbe0.5%
      Radius/fr2
      Thickness/mm0.02
      Decenter (surface and element)/mm0.01
      Tilt (surface and element)/(')0.5
      Irregularity0.25
    Tools

    Get Citation

    Copy Citation Text

    Zhikun Wu, Entao Shi, Yongmei Wang. Design of PGP imaging spectrometer with eliminating spectral line bending[J]. Infrared and Laser Engineering, 2021, 50(6): 20200433

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical design

    Received: Nov. 6, 2020

    Accepted: --

    Published Online: Aug. 19, 2021

    The Author Email: Entao Shi (set@nssc.ac.cn)

    DOI:10.3788/IRLA20200433

    Topics