Acta Photonica Sinica, Volume. 49, Issue 6, 0630001(2020)

Experimental Study of Radiometer Based on Spontaneous Parametric Down-conversion Calibration

You-bo HU1,2, Dong-yang GAO3, Jian-jun LI1, and Xiao-bing ZHENG1、*
Author Affiliations
  • 1Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China
  • 2University of Sciences and Technology of China, Hefei 230026, China
  • 3Shanghai Aerospace Control Technology Research Institute, Shanghai 201109, China
  • show less
    Figures & Tables(11)
    Diagram of the direction of photon emission generated by SPDC
    The relationship between wavelength and divergence angle of SPDC
    Principle of absolute quantum efficiency calibration for detectors with loss or decay
    Schematic diagram of the radiometer
    Schematic diagram of spontaneous parametric down conversion spectrum measurement
    Experimental diagram of radiometer loss or decay test
    Photon ring images with different wavelengths
    Coincidence peaks with different wavelengths
    • Table 1. Theoretical value and measured value of divergence angle

      View table
      View in Article

      Table 1. Theoretical value and measured value of divergence angle

      WavelengthDivergence angleTheoretical value
      466nm5.305.6
      550nm4.144.2
      580nm3.483.6
      605nm2.982.8
      685nm1.000.6
      737nm1.230.4
      850nm3.603.1
      910nm5.305.1
      1000nm7.507.2
    • Table 2. Radiance of halogen-tungsten lamp measured by radiometer under different optical decay conditions

      View table
      View in Article

      Table 2. Radiance of halogen-tungsten lamp measured by radiometer under different optical decay conditions

      NumberPhoton counting rate in the observation mode/sChannel detection efficiencyThe radiance of output port/(W·sr-1·mm-2)
      12 368 3210.086 59.27×10-9
      21 846 3240.067 79.22×10-9
      31 456 8750.053 49.18×10-9
      41 267 3160.046 39.22×10-9
      5954 3420.034 79.19×10-9
      Relatively standard uncertainty0.38%
    • Table 3. Radiance of halogen-tungsten lamp measured by radiometer under different electronics decay conditions

      View table
      View in Article

      Table 3. Radiance of halogen-tungsten lamp measured by radiometer under different electronics decay conditions

      NumberPhoton counting rate in the observation mode/sChannel detection efficiencyThe radiance of output port/(W·sr-1·mm-2)
      12 717 6550.029 4851.202×10-8
      22 696 4840.029 2171.203×10-8
      3469 5310.004 6811.208×10-8
      Relatively standard uncertainty0.28%
    Tools

    Get Citation

    Copy Citation Text

    You-bo HU, Dong-yang GAO, Jian-jun LI, Xiao-bing ZHENG. Experimental Study of Radiometer Based on Spontaneous Parametric Down-conversion Calibration[J]. Acta Photonica Sinica, 2020, 49(6): 0630001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Spectroscopy

    Received: Dec. 19, 2019

    Accepted: Mar. 12, 2020

    Published Online: Nov. 26, 2020

    The Author Email: ZHENG Xiao-bing (xbzheng@aiofm.ac.cn)

    DOI:10.3788/gzxb20204906.0630001

    Topics