Infrared and Laser Engineering, Volume. 51, Issue 12, 20220246(2022)

Opto-mechanical design and performance test of infrared channel field radiometer

Yunxiang Zhang1, Xin Li1、*, Dong Huang1,2, Yanna Zhang1, Wei Wei1, and Yan Pan1,2
Author Affiliations
  • 1Key Laboratory of Optical Calibration and Characterization, Anhui Institute of Optics and Fine Mechanics,Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, China
  • 2University of Science and Technology of China, Hefei 230026, China
  • show less
    Figures & Tables(21)
    Schematic of ICFR
    Schematic diagram of combined optical path
    Optical system of ICFR
    Diffuse spots of optical system of ICFR
    Schematic diagram of ICFR optical head structure
    Light barrel structure diagram of ICFR
    Radiance calibration curve of the ICFR four channels
    Schematic diagram of thermal shock resistance measuring device
    Temperature setting of Peltier
    Difference between ICRF measured blackbody temperature and set temperature under thermal shock
    Working environment temperature suitability test device of ICFR
    Result of ICFR measures the brightness temperature of the reference plate and Platinum resistance measures the physical temperature of the reference plate
    Field comparison experiment
    Comparison of the surface brightness temperature data obtained by the corresponding channels of ICFR and CE312
    • Table 1. Technical indicators of ICFR

      View table
      View in Article

      Table 1. Technical indicators of ICFR

      ParametersValues
      Wavelength/μm8-14, 8.2-9.2, 10.3-11.3, 11.5-12.5
      Field of view/(°)10
      Ambient temperature/℃−20-50
      Temperature resolution/K0.01
      Uncertainty of radiometric calibration/K0.16
    • Table 2. Lens characteristics of ICFR

      View table
      View in Article

      Table 2. Lens characteristics of ICFR

      LensFocal length/mmEffective diameter/mm
      Objective lens25.45.6
      Field lens3.234.35
    • Table 3. Characteristics of ST60 TO-5

      View table
      View in Article

      Table 3. Characteristics of ST60 TO-5

      ParametersValues
      Active area size/mm0.61×0.61
      Number of junctions80
      Signal-to-noise ratio/Hz1/24792
      Responsivity/V·W−1105.9
      Resistance/kΩ105.9
      Temperature coefficient of R/0.11%
      Noise voltage/nV·Hz1/233.8
      Noise equivalent powe/nW·Hz1/20.52
      Detectivity/cm−1·Hz1/2·W−12.38×108
      Time constant/ms18
      Field of view/(°)81
      Germanium window with nonreflective coating/μm 8–14
    • Table 4. Performance parameters of CES100-06 blackbody

      View table
      View in Article

      Table 4. Performance parameters of CES100-06 blackbody

      ParameterValue
      Radiation area/mm2250×250
      Temperature range/℃−50-100
      Uniformity±0.03% thereafter within central 80%
      Temperature control accuracy/℃±0.03, 0-100 ±0.05, −50-0
      Emissivity0.992±0.001
      Controller size/mm133.35×482.6×457.2
      Voltage VAC100-240±10% (50/60 Hz)
    • Table 5. Values of the parameters fitting the ICFR channels

      View table
      View in Article

      Table 5. Values of the parameters fitting the ICFR channels

      Channel/μmaibinidi
      8-14989.142281303.313970.986013.29853
      8.2-9.22724.557311629.113140.992391.73586
      10.3-11.3781.956131307.124260.99781.02869
      11.5-12.5452.369751153.004130.994860.98921
    • Table 6. Combined uncertain of radiometric calibration

      View table
      View in Article

      Table 6. Combined uncertain of radiometric calibration

      Sources of uncertaintySymbolUncertainty typeUncertainty
      Blackbody emissivityu1B0.100%
      Blackbody temperature accuracyu2B0.043%
      Calibration coefficient fittingu3A0.019%
      Uniformityu4B0.03%
      Combined uncertaintyuc-0.114%
      Equivalent temperature uncertainty (K@303 K, 11 μm)--0.16(k=2)
    • Table 7. Analysis of the mean value and standard deviation of the brightness temperature of the ground surface measured by ICFR and CE312

      View table
      View in Article

      Table 7. Analysis of the mean value and standard deviation of the brightness temperature of the ground surface measured by ICFR and CE312

      Bands/μmMean/KStd dev/KMin/KMax/K
      8.2-9.20.06510.26134−0.843150.86688
      10.2-11.20.090890.2104−0.535710.68209
      11.5-12.50.035260.30211−0.928530.9083
      8-14−0.07450.18278−0.970530.5863
    Tools

    Get Citation

    Copy Citation Text

    Yunxiang Zhang, Xin Li, Dong Huang, Yanna Zhang, Wei Wei, Yan Pan. Opto-mechanical design and performance test of infrared channel field radiometer[J]. Infrared and Laser Engineering, 2022, 51(12): 20220246

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Infrared technology and application

    Received: May. 10, 2022

    Accepted: --

    Published Online: Jan. 10, 2023

    The Author Email: Li Xin (xli@aiofm.ac.cn)

    DOI:10.3788/IRLA20220246