Acta Optica Sinica, Volume. 42, Issue 11, 1134023(2022)

Filtered Back-Projection Reconstruction for Wide Field of View Micro-Computed Tomography Based on Array Microfocus X-Ray Source

Dongyang Xie1,2, Yuchuan Liu1,2, Fengxiao Li1,2, Yuanzhi Tian1,2, Liping Tang1,2, and Rifeng Zhou1,2、*
Author Affiliations
  • 1Key Lab of Optoelectronic Technology and Systems, Ministry of Education, Chongqing University, Chongqing 400044, China
  • 2Engineering Research Center of Industrial Computed Tomography Nondestructive Testing, Ministry of Education, Chongqing University, Chongqing 400044, China
  • show less
    Figures & Tables(12)
    Array microfocal ray source. (a) Physical photo; (b) structural diagram; (c) projections of array microfocal points; (d) sample
    Schematic diagram of EBMCT imaging model. (a) Scanning mode; (b) definition of geometric parameters
    Smooth weighting strategy. (a) Projection data redundancy of adjacent focal spots; (b) weighting function curves
    Weighting process of projection data. (a) Projection data before weighting processing; (b) visualized weighting value; (c) projection data after weighting process
    Simulation results. (a) Shepp-Logan model; (b) reconstructed image of FBMCT; reconstructed images of EBMCT (c) without and (d) with weighting processing; (e) profile of horizontal center row
    Comparison of algorithms. (a) Reconstructed image of FORBILD model; reconstructed images of (b) rebinning algorithm, (c) SART algorithm, and (d) MSFBP algorithm; (e)--(h) reconstructed images of ROI of Figs. 6(a)--(d)
    Prototype of EBMCT
    Reconstructed images of two kinds of CT. (a) Image reconstructed by FBMCT (field of view is 3.8 mm); (b) image reconstructed by EBMCT (field of view is 9.5 mm); (c) sample photo
    Reconstructed images of ROI obtained by three algorithms. (a) Rebinning algorithm; (b) SART algorithm; (c) MSFBP algorithm
    Test experiments of MTF. (a) Image of pin gauge reconstruction; (b) MTF curves
    • Table 1. Scanning parameters of simulation

      View table

      Table 1. Scanning parameters of simulation

      ParameterEBMCTFBMCT
      Number of source spots51
      Width of source array /mm6-
      Distance from source arrayto detector /mm300300
      Distance from source arrayto isocenter /mm1515
      Diameter of FOV /mm9.53.8
      Number of views360360
      Width of detector elements /μm74.874.8
      Number of detector elements10241024
    • Table 2. Evaluation of reconstructed image quality

      View table

      Table 2. Evaluation of reconstructed image quality

      IndicatorRebinningalgorithmSARTalgorithmMSFBPalgorithm
      RMSE0.36330.21460.2173
      PSNR13.900918.472518.3657
      SSIM0.89110.96750.9663
    Tools

    Get Citation

    Copy Citation Text

    Dongyang Xie, Yuchuan Liu, Fengxiao Li, Yuanzhi Tian, Liping Tang, Rifeng Zhou. Filtered Back-Projection Reconstruction for Wide Field of View Micro-Computed Tomography Based on Array Microfocus X-Ray Source[J]. Acta Optica Sinica, 2022, 42(11): 1134023

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: X-Ray Optics

    Received: Dec. 23, 2021

    Accepted: Feb. 9, 2022

    Published Online: Jun. 3, 2022

    The Author Email: Zhou Rifeng (cquictzhourf@163.com)

    DOI:10.3788/AOS202242.1134023