Chinese Journal of Lasers, Volume. 51, Issue 3, 0307401(2024)

Lensless Refractive Measurement System Based on Shack‐Hartmann Wavefront Detection

Kangjie Geng1,2, Hetong Zhang1,2, Shangshang Ding1,2, Yang Zhang1,2, Min Liu1,2、*, and Weiwei Fu1,2、**
Author Affiliations
  • 1Division of Life Sciences and Medicine, School of Biomedical Engineering (Suzhou), University of Science and Technology of China, Hefei 230000, Anhui , China
  • 2Suzhou Institute of Biomedical Engineering and Technology, Chinese Academy of Sciences, Suzhou 215163, Jiangsu , China
  • show less
    Figures & Tables(11)
    Principle of Schack-Hartmann wavefront sensor
    Relationship between spot offset and wavefront slope
    Modeling of the image acquisition optical path
    Hartmann photogram
    Actual measured diopter as a function of distance between Schack-Hartmann sensor and human eye
    Schematic illustration of the variation of wavefront with propagation distance in myopic eyes
    True diopter corrected by the distal point formula
    Optical schematic
    Physical drawing of the test prototype
    • Table 1. Parameters of the eye model

      View table

      Table 1. Parameters of the eye model

      Ocular structureRadius of curvature-front surface /mmRadius of curvature-rear surface /mmThicknesses /mmIndex of refraction
      Retina-12.3-12.3
      Vitreous humor16.601.3360
      Lens10.26.04.001.4200
      Aqueous humor3.051.3374
      Cornea7.86.50.551.3771
    • Table 2. Measurement results

      View table

      Table 2. Measurement results

      Eye model diopter /DSpherical diopterCylindrical diopter
      Average value /DCoefficient of variationMaximum error /DAverage value /DCoefficient of variationMaximum error /D
      -109.9997470.0067760.14-0.2295950.0873570.24
      -55.0002050.0079380.07-0.0182360.0819480.20
      -2.52.4762660.0155440.06-0.1504830.0685700.15
      00.0009130.06-0.1103230.0835790.12
      2.5-2.4918960.0295750.09-0.1394990.0616460.15
      5-5.0001080.0276960.16-0.2341970.0256860.15
      10-9.9998570.0233530.20-0.1516670.0759140.23
    Tools

    Get Citation

    Copy Citation Text

    Kangjie Geng, Hetong Zhang, Shangshang Ding, Yang Zhang, Min Liu, Weiwei Fu. Lensless Refractive Measurement System Based on Shack‐Hartmann Wavefront Detection[J]. Chinese Journal of Lasers, 2024, 51(3): 0307401

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Bio-Optical Sensing and Manipulation

    Received: Sep. 24, 2023

    Accepted: Nov. 16, 2023

    Published Online: Jan. 24, 2024

    The Author Email: Liu Min (lium@sibet.ac.cn), Fu Weiwei (fuww@sibet.ac.cn)

    DOI:10.3788/CJL231325

    Topics