    Figures & Tables(8)
    Flow chart of sorting
    Volume frequency distribution curve of simulation results
    Volume frequency distribution curve of experimental results
    • Table 1. Simulation results of unimodal distribution

      View table
      View in Article

      Table 1. Simulation results of unimodal distribution

      设定值$ \left( {\overline D ,k} \right) $反演值$ \left( {\overline D ,k} \right) $R
      (8.155,9.349)(8.16,9.28)0.0076
      (12.734,3.352)(12.9,3.24)0.0412
    • Table 2. Simulation results of bimodal distribution

      View table
      View in Article

      Table 2. Simulation results of bimodal distribution

      设定值$ \left(\overline{{D}_{1}},{k}_{1},n,\overline{{D}_{2}},{k}_{2}\right) $反演值$ \left(\overline{{D}_{1}},{k}_{1},n,\overline{{D}_{2}},{k}_{2}\right) $RRMSE
      (3.358,11.71,0.5,10.242,11.452)(3.36,11.65,0.5,10.24,11.50)0.0021
      (3.164,11.52,0.7,10.195,11.853)(3.16,11.85,0.7,10.14,11.75)0.0638
    • Table 3. Simulation results after adding noise

      View table
      View in Article

      Table 3. Simulation results after adding noise

      设定值随机噪声反演值R
      ( 8.155, 9.349)1%( 8.199, 9.242)0.0351
      3%( 8.24, 9.484)0.0646
      (3.358,11.71,0.5,10.242,11.452)1%(3.3880,11.3100,0.5000,10.1440,11.2650)0.0407
      3%(3.3420,11.6500,0.5000,10.3900,11.7950)0.0612
    • Table 4. Experimental results of unimodal distribution of standard particles

      View table
      View in Article

      Table 4. Experimental results of unimodal distribution of standard particles

      样本编号D50标称值/${\text{μm} }$D50实测值/${\text{μm} }$相对误差
      GBW1201343.443.7180.0808
      GBW1200249.409.6610.0278
      GBW12004114.6215.680.0725
    • Table 5. Experimental results of bimodal distribution of standard particles

      View table
      View in Article

      Table 5. Experimental results of bimodal distribution of standard particles

      标称值 D50/μm实测值 D50/μm相对误差
      (3.44,9.40)(3.201,10.180)(0.0695,0.0830)
    Tools

    Get Citation

    Copy Citation Text

    Hongqin WANG, Zexi ZHENG, Huazhong XIANG, Xiangpeng LI. Particle size measurement by near-infrared light extinction method using database technologies[J]. Optical Instruments, 2023, 45(6): 1

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Jan. 30, 2023

    Accepted: --

    Published Online: Feb. 29, 2024

    The Author Email: ZHENG Zexi (zexizheng@outlook.com)

    DOI:10.3969/j.issn.1005-5630.202301300006

    Topics