Spectroscopy and Spectral Analysis, Volume. 42, Issue 2, 622(2022)

Determination of Trace Oxygen in High Purity Gold by Using Inert Gas Fusion Infrared Spectroscopy

Yi-chuan TANG1、*, Yan-jie CUI1、1;, Jian-ying ZHANG1、1;, Sheng HE2、2;, Tao ZHOU1、1; *;, and Bing WU1、1;
Author Affiliations
  • 11. National Institute of Metrology, China, Beijing 100029, China
  • 22. Beijing Research Institute of Uranium Geology, Beijing 100029, China
  • show less
    Figures & Tables(7)
    Results of Oxygen in gold(a): 1st measurement of gold; (b): 2nd measurement of gold;(c): measurement of gold and tin
    Trend of results of oxygen in gold with different mass of tin particle
    Linear calibration curve of oxygen
    • Table 1. Results of oxygen blank in tin particle (mg·kg-1, n=12)

      View table
      View in Article

      Table 1. Results of oxygen blank in tin particle (mg·kg-1, n=12)

      No.ResultsNo.Results
      10.06870.079
      20.07180.060
      30.08090.079
      40.050100.067
      50.080110.066
      60.083120.067
      平均值0.07
      标准偏差0.01
      定量限0.1
    • Table 2. Measurement results of oxygen in high purity gold

      View table
      View in Article

      Table 2. Measurement results of oxygen in high purity gold

      No.氧含量测量值
      /(mg·kg-1)
      No.氧含量测量值
      /(mg·kg-1)
      10.98470.957
      21.09681.117
      31.10991.224
      40.905101.103
      51.170111.058
      60.987121.327
      AVE1.09
      RSD/%12
      校正值1.1
    • Table 3. Measurement results of oxygen in high purity gold by using addition method

      View table
      View in Article

      Table 3. Measurement results of oxygen in high purity gold by using addition method

      No.金质量/g加标量/μg测得总量/μg回收率/%
      10.49814.85417.681103
      20.5044.3026.66899
      30.50547.59652.635105
      40.5129.96312.30499
      50.50137.08939.737101
    • Table 4. Measurement results of oxygen in high purity gold by using SIMS method

      View table
      View in Article

      Table 4. Measurement results of oxygen in high purity gold by using SIMS method

      No.SRM685w
      /cps
      High purity
      gold/cps
      High purity gold
      /(mg·kg-1)
      1279 074152 4831.10
      2321 085136 7960.98
      3183 148144 0831.04
      4338 216133 0160.96
      5356 608114 4020.82
      6234 436113 0900.81
      7298 16483 7180.60
      8241 209150 7881.08
      9268 043132 5030.95
      10250 66094 8140.68
      11286 967111 3750.80
      12281 598152 4831.10
      AVE278 267126 6290.91
      RSD/%171818
    Tools

    Get Citation

    Copy Citation Text

    Yi-chuan TANG, Yan-jie CUI, Jian-ying ZHANG, Sheng HE, Tao ZHOU, Bing WU. Determination of Trace Oxygen in High Purity Gold by Using Inert Gas Fusion Infrared Spectroscopy[J]. Spectroscopy and Spectral Analysis, 2022, 42(2): 622

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Research Articles

    Received: Jan. 13, 2021

    Accepted: Feb. 24, 2021

    Published Online: Apr. 2, 2022

    The Author Email: TANG Yi-chuan (tangych@nim.ac.cn)

    DOI:10.3964/j.issn.1000-0593(2022)02-0622-05

    Topics