Acta Photonica Sinica, Volume. 51, Issue 4, 0412005(2022)

Geometric Parameters Calibration Method for Optically CCD-coupled CT System

Jing ZOU... Yiming JIANG, Xiaodong HU and Jintao ZHAO* |Show fewer author(s)
Author Affiliations
  • School of Precision Instrument and Opto-electronics Engineering,Tianjin University,Tianjin 300072,China
  • show less
    Figures & Tables(9)
    The structure diagram of CCD-coupled CT system
    Diagram of coordinate system establishment
    Projection diagram
    Experimental device
    Extraction center
    Comparison of bamboo fiber reconstruction results
    • Table 1. Experiment condition

      View table
      View in Article

      Table 1. Experiment condition

      ParametersValue
      SOD/mm5
      SDD/mm10
      Target power/W0.8
      Current/μA400
      Exposure time/s15
      Frame number360
    • Table 2. Calibration results of the parameters

      View table
      View in Article

      Table 2. Calibration results of the parameters

      ParametersValue
      η/(°)1.32
      R/mm4.443
      D/mm11.824
      u0/μm81
      v0/μm-165
    • Table 3. Quantitative evaluation of EOG

      View table
      View in Article

      Table 3. Quantitative evaluation of EOG

      EOG
      Before calibration8.04×108
      After calibration9.57×108
    Tools

    Get Citation

    Copy Citation Text

    Jing ZOU, Yiming JIANG, Xiaodong HU, Jintao ZHAO. Geometric Parameters Calibration Method for Optically CCD-coupled CT System[J]. Acta Photonica Sinica, 2022, 51(4): 0412005

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Oct. 18, 2021

    Accepted: Nov. 24, 2021

    Published Online: May. 18, 2022

    The Author Email: ZHAO Jintao (zjt2019@tju.edu.cn)

    DOI:10.3788/gzxb20225104.0412005

    Topics