Infrared and Laser Engineering, Volume. 54, Issue 1, 20240221(2025)

Preliminary study on test method and life prediction of single failure mechanism of rotary Stirling cooler

Haiying LI, Kai ZHANG, Yun ZHA, Chonghua JIANG, Jian HUAN, and Wenli ZHAO
Author Affiliations
  • Kunming Institute of Physics, Kunming 650223, China
  • show less
    Figures & Tables(6)
    Outline of metal C seal ring
    Compression-load curves of metal C seal ring
    Microstructure diagram of coating interface (Sample No.3)
    Profile topography of sealed area (failure sample)
    • Table 1. Description of sample 80 ℃ storage test

      View table
      View in Article

      Table 1. Description of sample 80 ℃ storage test

      SamplenumberUsage during storageStorage time/hDescription of test
      1a) The sample has carried out product acceptance test;b) The average storage temperature of the samples in the field is about 20 ℃.5760The leakage rate, cooling time and power consumption of each periodic test sample are all within the qualified range.
      2a) The sample has carried out product acceptance test;b) The samples are mainly used for intermittent commissioning in the laboratory, and the time is about 1 to 2 years;c) The samples will be placed on the laboratory shelf in the later stage at an average ambient temperature of about 23 ℃.1440After 1440 h storage,the leakage rate of the sample is normal and the dewar vacuum fails. After separation, the cooling time and power consumption of the cooler are tested by the test dewar.
    • Table 2. Description of the temperature cycle test of sample under the −55-80 ℃

      View table
      View in Article

      Table 2. Description of the temperature cycle test of sample under the −55-80 ℃

      SamplenumberUsage during storageNumber of cycles/timesDescription of test
      3a) The sample has carried out product acceptance test;b) The average temperature difference between high and low in the storage environment of the samples is about 10 ℃, and the highest temperature is about 30 ℃.330After the 330th time, the sample fails due to excessive leakage rate.
      4a) The sample has carried out product acceptance test;b) The samples are mainly used for intermittent commissioning in the laboratory, and the time is about 1 to 2 years;c) The samples will be placed on the laboratory shelf in the later stage, with the average temperature difference between high and low about 5 ℃ and the highest temperature about 28 ℃.270After the 270th time, the sample fails due to excessive leakage rate.
    Tools

    Get Citation

    Copy Citation Text

    Haiying LI, Kai ZHANG, Yun ZHA, Chonghua JIANG, Jian HUAN, Wenli ZHAO. Preliminary study on test method and life prediction of single failure mechanism of rotary Stirling cooler[J]. Infrared and Laser Engineering, 2025, 54(1): 20240221

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: 红外技术及应用

    Received: Oct. 27, 2024

    Accepted: --

    Published Online: Feb. 12, 2025

    The Author Email:

    DOI:10.3788/IRLA20240221

    Topics