Optical Instruments, Volume. 46, Issue 3, 40(2024)

Discussion on quality control of liver function reserve analyzer (ICG)

Xin JIN1... Luping WU2 and Tingting ZHU1,* |Show fewer author(s)
Author Affiliations
  • 1Shanghai Seventh People's Hospital, Shanghai 200137, China
  • 2School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
  • show less
    Figures & Tables(6)
    Principle of liver function reserve analyzer
    Light-emitting diode
    Photodiode
    Experimental related equipment and reagents
    Test results
    • Table 1. Average detection test results

      View table
      View in Article

      Table 1. Average detection test results

      名称试验1试验2试验3试验4试验5试验6
      标准浓度100%80%60%40%20%10%
      探头1100%80%59.9%40%19.8%10.1%
      探头2100%79.7%59.3%39.5%19.6%9.4%
      探头3100%79.1%58.9%39.3%18.1%8.8%
      探头498.7%78.5%57.2%37.6%17.4%7.2%
    Tools

    Get Citation

    Copy Citation Text

    Xin JIN, Luping WU, Tingting ZHU. Discussion on quality control of liver function reserve analyzer (ICG)[J]. Optical Instruments, 2024, 46(3): 40

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Feb. 22, 2024

    Accepted: --

    Published Online: Jul. 31, 2024

    The Author Email: ZHU Tingting (skyyagami97@163.com)

    DOI:10.3969/j.issn.1005-5630.202402220023

    Topics