Laser & Optoelectronics Progress, Volume. 58, Issue 3, 3120041(2021)

Three-Channel Frequency-Modulated Continuous-Wave Laser Interferometric Displacement Measurement System

Sheng Qiming1, Zheng Gang2、*, Zhang Xiongxing2, Sun Bin2, Jing Liqiang2, and Han Yuan1
Author Affiliations
  • 1School of Electronic Information Engineering, Xi′an Technological University, Xi′an, Shaanxi 710021, China
  • 2School of Optoelectronic Engineering, Xi′an Technological University, Xi′an, Shaanxi 710021, China
  • show less
    Figures & Tables(14)
    Angular frequency waveforms of different signals
    Schematic of three-channel FMCW laser interference system
    Control circuit of three-channel FMCW laser interferometer
    Schematic of laser drive circuit
    Circuit board for three-channel FMCW laser interference system
    Execution sequence of each chip program
    Flowchart of phase detection of beat signal
    Experimental device
    Beat signal before and after adding synchronization signal. (a) Before adding; (b) after adding
    Correlation of feature points before and after adding synchronization signal. (a) Before adding; (b) after adding
    Scatter points of displacement errors of different channels. (a) Channel 1; (b) channel 2; (c) channel 3
    Linearity curves of three channels
    • Table 1. Routing length of each chip signal line

      View table

      Table 1. Routing length of each chip signal line

      MCUWire length of clock /mmWire length of synchronous signal /mm
      MCU 149.5290
      MCU 249.53522.276
      MCU 349.53122.294
    • Table 2. Stability test results

      View table

      Table 2. Stability test results

      Number of experimentsMean/nmStandard deviation/nm
      1-0.2896502.64858
      20.2655501.79282
      -0.6098502.52146
      3

      4

      5

      6

      -0.1635751.75456
      0.2361302.15621
      0.5963302.02113
      Average0.3601802.14913
    Tools

    Get Citation

    Copy Citation Text

    Sheng Qiming, Zheng Gang, Zhang Xiongxing, Sun Bin, Jing Liqiang, Han Yuan. Three-Channel Frequency-Modulated Continuous-Wave Laser Interferometric Displacement Measurement System[J]. Laser & Optoelectronics Progress, 2021, 58(3): 3120041

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Jun. 1, 2020

    Accepted: --

    Published Online: Mar. 12, 2021

    The Author Email: Gang Zheng (zhengg@xatu.edu.cn)

    DOI:10.3788/LOP202158.0312004