Chinese Physics B, Volume. 29, Issue 9, (2020)

Fabrication and performance evaluation of GaN thermal neutron detectors with 6LiF conversion layer

Zhifu Zhu1...2,3, Zhijia Sun2,†, Jijun Zou3, Bin Tang2, Qinglei Xiu2, Renbo Wang3, Jinhui Qu3, Wenjuan Deng3, Shaotang Wang3, Junbo Peng3, Zhidong Wang3, Bin Tang3 and Haiping Zhang4 |Show fewer author(s)
Author Affiliations
  • 1State Key Laboratory Breeding Base of Nuclear Resources and Environment, East China University of Technology, Nanchang 33003, China
  • 2State Key Laboratory of Particle Detection and Electronics, Institute of High Energy Physics, Chinese Academy of Sciences, Beijing 100049, China
  • 3Engineering Research Center of Nuclear Technology Application (East China University of Technology), Ministry of Education, Nanchang 001, China
  • 4CGN Begood Technology Co., Ltd., Nanchang 330013, China
  • show less
    Figures & Tables(7)
    (a) Device structure evaporated 6LiF; (b) photograph of nine neutron detectors mounted in the PCB with Au-plated contact by 25-μm gold wires. External ring contact (Ti/Al/Ni/Au) and internal circular contact (Ni/Au) were connected to the VCC and GND of the applied voltage, respectively. The dimensions in the picture are not drawn to scale.
    Neutron detection-system diagram using ORTEC modular pulse-processing electronic analyzer.
    The thermal neutron capture yield produced by fast neutrons captured with PE as a function of the thicknesses using GENT4.
    Thermal neutron detection efficiency as a function of the 6LiF converter thickness.
    The I–V characteristics of neutron detectors before and after deposition of 6LiF under various reverse bias voltages.
    Pulse-height spectrum of neutron detector irradiated by 5.48-MeV 241Am alpha particles as a function of reverse bias voltages at room temperature.
    • Table 1. Pulse counts of neutron detector from pulse-counting board under different biases voltages.

      View table
      View in Article

      Table 1. Pulse counts of neutron detector from pulse-counting board under different biases voltages.

      Voltage/VPulse counts (t = 1 h)Total countsThreshold voltage/mV
      123456
      024312731282716850
      5202622323224156100
      10192820332129150150
    Tools

    Get Citation

    Copy Citation Text

    Zhifu Zhu, Zhijia Sun, Jijun Zou, Bin Tang, Qinglei Xiu, Renbo Wang, Jinhui Qu, Wenjuan Deng, Shaotang Wang, Junbo Peng, Zhidong Wang, Bin Tang, Haiping Zhang. Fabrication and performance evaluation of GaN thermal neutron detectors with 6LiF conversion layer[J]. Chinese Physics B, 2020, 29(9):

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: May. 3, 2020

    Accepted: --

    Published Online: Apr. 29, 2021

    The Author Email: Sun Zhijia (jjzou@ecit.cn)

    DOI:10.1088/1674-1056/ab9c05

    Topics