Acta Optica Sinica, Volume. 38, Issue 6, 0612002(2018)
Accuracy in Refractive Index Measurement of As2Se3 Chalcogenide Glass by IR Spectroscopic Ellipsometer
The As2Se3 specimens with different thicknesses, roughnesses and surface smoothnesses are prepared by use of homemade As2Se3 glass rods. The refractive indices are measured by the infrared spectroscopic ellipsometer, and an optical model is built to obtain the refractive indices by fitting. The influences of thickness, roughness and surface smoothness on the refractive indices of specimens are comparatively analyzed. The results show that all factors have obvious influences on the measurement accuracy of the ellipsometer and the surface smoothness is of most importance. If the specimen thickness is controlled to be 1-3 mm, and simultaneously the back side roughness and surface smoothness of specimens are increased, the measurement accuracy of the ellipsometer can be obviously enhanced.
Get Citation
Copy Citation Text
Yang Li, Yongxing Liu, Shixun Dai, Tiefeng Xu, Changgui Lin, Feifei Chen. Accuracy in Refractive Index Measurement of As2Se3 Chalcogenide Glass by IR Spectroscopic Ellipsometer[J]. Acta Optica Sinica, 2018, 38(6): 0612002
Category: Instrumentation, Measurement and Metrology
Received: Oct. 20, 2017
Accepted: --
Published Online: Jul. 9, 2018
The Author Email: Xu Tiefeng (xutiefeng@nbu.edu.cn)