Spectroscopy and Spectral Analysis, Volume. 33, Issue 8, 2039(2013)

Effect of Flat-Mirror Device on Laser-Induced Plasma Radiation Characteristics

CHEN Jin-zhong*, BAI Jin-ning, CHEN Zhen-yu, CHENG Chen, SUN Jiang, and WEI Yan-hong
Author Affiliations
  • [in Chinese]
  • show less

    To improve the quality of laser-induced breakdown spectroscopy, flat-mirror device was proposed. The effects of flat-mirror device on the radiation characteristics of laser-induced plasma were studied. The experimental results showed that when the device consisted of three flat-mirrors placed around the plasma, the spectral line intensity of Mg, Fe, Ba, Ti and Al increases by about 116.2%, 96.43%, 90.93%, 102.1% and 98.57% than that without flat-mirror device, and the signal-to-noise raises by around 39.17%, 32.48%, 38.07%, 39.95% and 21.30%, respectively. By measuring the plasma parameters, the mechanism of the radiation enhancement obtained with the device consisting of three flat-mirrors was explained. This method was an effective way to improve the detection capacity of LIBS.

    Tools

    Get Citation

    Copy Citation Text

    CHEN Jin-zhong, BAI Jin-ning, CHEN Zhen-yu, CHENG Chen, SUN Jiang, WEI Yan-hong. Effect of Flat-Mirror Device on Laser-Induced Plasma Radiation Characteristics[J]. Spectroscopy and Spectral Analysis, 2013, 33(8): 2039

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Dec. 11, 2012

    Accepted: --

    Published Online: Aug. 12, 2013

    The Author Email: Jin-zhong CHEN (chenjinzhongcn@126.com)

    DOI:10.3964/j.issn.1000-0593(2013)08-2039-04

    Topics