Acta Optica Sinica, Volume. 29, Issue 5, 1275(2009)

Measurement Method of X-ray Focus Projection Coordinates of Three-Dimensional Micro-Computerized Tomography Scanning System

Yang Min*, Liu Jinghua, Li Baolei, Wu Wenjin, and Wang Gang
Author Affiliations
  • [in Chinese]
  • show less

    In the three-dimensional micro-computerized tomography scanning system, it is impossible to measure the positions of the X-ray focus and imaging plane of the detector by direct means. A simple and feasible method to measure the X-ray focus projection coordinates with high accuracy is proposed. This method is based on the theory that the projection of the sphere object in the cone-beam field is an ellipse. Through acquiring the digital radiography images of the double sphere objects, the image and graphics processing methods and the least square fitting method are used to get the intersection point of the major axis of the two ellipses. The intersection point is also the X-ray focus projection on imaging plane of the detector. The experimental results prove that the accuracy of this method can satisfy the requirements of the practical micro-comptuterized tomography scanning system, and is easy to realize with high anti-noise ability.

    Tools

    Get Citation

    Copy Citation Text

    Yang Min, Liu Jinghua, Li Baolei, Wu Wenjin, Wang Gang. Measurement Method of X-ray Focus Projection Coordinates of Three-Dimensional Micro-Computerized Tomography Scanning System[J]. Acta Optica Sinica, 2009, 29(5): 1275

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Nov. 28, 2008

    Accepted: --

    Published Online: May. 22, 2009

    The Author Email: Min Yang (minyang.ndt@263.net)

    DOI:

    Topics