Chinese Optics Letters, Volume. 12, Issue 3, 031202(2014)

A three-spectrum method for accurate calibration of bi-plate zero-order retarder

Zhigang Han, Zhi Xu, and Lei Chen

A new spectroscopic method is proposed for the characterization of optical zero-order retarders. It is demonstrated that the retardance as well as the variation of the effective fast axis of a bi-plate zero-order quarter retarder (633 nm) can be obtained with high accuracy in a broadband wavelength range by taking spectra at only three independent angular orientations of the retarder. The calibration results excellently agree with theoretical models, indicating the new method could be used as a simple and reliable way for efficient self-spectral-calibration of optical zero-order retarders.

Tools

Get Citation

Copy Citation Text

Zhigang Han, Zhi Xu, Lei Chen. A three-spectrum method for accurate calibration of bi-plate zero-order retarder[J]. Chinese Optics Letters, 2014, 12(3): 031202

Download Citation

EndNote(RIS)BibTexPlain Text
Save article for my favorites
Paper Information

Category: Instrumentation, measurement, and metrology

Received: Dec. 12, 2013

Accepted: Jan. 15, 2014

Published Online: Mar. 17, 2014

The Author Email:

DOI:10.3788/col201412.031202

Topics