Chinese Journal of Lasers, Volume. 42, Issue 5, 516003(2015)
Fiber Spectrometer Optical Simulation Optimization and Calibration
The optical system is the core part of a spectrometer, and it determines the basic properties and the volume of the entire instrument. Advantages and disadvantages of several optical systems are analyzed, and the asymmetric Czerny-Turner system is adopted. The resulting optical path optimized with Zemax has an overall resolution of 1.8 nm, and a measurement range of 200~900 nm. Relative error of wavelength is controlled to be below 0.05 nm, with cubic spline interporlation coefficients obtained from Matlab and HG- 1 mercury- argon used as the calibration light source. Analyses on data for comparision with the USB4000 fiber spectrometer show that both of the design and implementation are feasible.
Get Citation
Copy Citation Text
Xu Danyang, Tong Jianping, Gao Jianxun, Wang Fei. Fiber Spectrometer Optical Simulation Optimization and Calibration[J]. Chinese Journal of Lasers, 2015, 42(5): 516003
Category: Optical Design and Fabrication
Received: Nov. 14, 2014
Accepted: --
Published Online: May. 6, 2015
The Author Email: Danyang Xu (xudanyang@zjut.edu.cn)