Spectroscopy and Spectral Analysis, Volume. 33, Issue 6, 1706(2013)

Optical Design of High-Resolution Echelle-Prism Cross-Dispersion

ZHANG Yin-xin1、*, YANG Huai-dong2, DENG Chao1, HUANG Zhan-hua1, and JIN Guo-fan2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    Aiming at the wide-spectrum high-resolution echelle spectrometer, the dispersion characteristic of the echelle grating and the cross-dispersion principle is discussed. And an optical design method of resolution preferential echelle-prism cross-dispersion is described, including the design of high-resolution main dispersion, the auxiliary dispersion for the separation on the overlapped orders, and the main-auxiliary dispersion calibration and revision. According to the requirements of commercial spectrometer, a design example is reported. The simulation and the experiment results show that when the spectral span is 400~900 nm, the resolution for Hg 546 nm is 51 000, and that for Na 589 nm is 44 000.

    Tools

    Get Citation

    Copy Citation Text

    ZHANG Yin-xin, YANG Huai-dong, DENG Chao, HUANG Zhan-hua, JIN Guo-fan. Optical Design of High-Resolution Echelle-Prism Cross-Dispersion[J]. Spectroscopy and Spectral Analysis, 2013, 33(6): 1706

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Sep. 19, 2012

    Accepted: --

    Published Online: Jun. 7, 2013

    The Author Email: Yin-xin ZHANG (yinxin@tju.edu.cn)

    DOI:10.3964/j.issn.1000-0593(2013)06-1706-05

    Topics