Acta Optica Sinica, Volume. 31, Issue s1, 100514(2011)
Design for Mounting Configuration of Ultra-Stable Fabry-Perot Cavity of Ultra-Narrow Linewidth Laser
Locking to the laser cavity length of ultra-stable cavities is an important method to gain ultra-short line width laser output. The stability of the laser relies on the length stability of the Fabry-Perot (F-P) cavities. Vibration induced cavity deformation is one of the dominate factors that affect the stability of ultra-stable optical cavities. The elastic deformation of two general cavities with different mounting configurations in vibration environment is quantitatively analyzed. Numerical results offer the optimum mounting configurations of the two cavities, where the change of vibration-induced cavity is minimized to 10-12 m
Get Citation
Copy Citation Text
Yang Tao, Li Ye, Gao Chunqing, Cao Jianping, Fang Zhanjun, Zang Erjun. Design for Mounting Configuration of Ultra-Stable Fabry-Perot Cavity of Ultra-Narrow Linewidth Laser[J]. Acta Optica Sinica, 2011, 31(s1): 100514
Category: OPTOELECTRONICS
Received: Mar. 1, 2011
Accepted: --
Published Online: Jun. 23, 2011
The Author Email: Tao Yang (tyang@bit.edu.cn)