Laser & Optoelectronics Progress, Volume. 57, Issue 5, 051202(2020)

Calibration Method for Stereo Phase Measuring Deflectometry System

Fangming Liu, Jiarui Lin, Yanbiao Sun, and Jigui Zhu*
Author Affiliations
  • State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China
  • show less

    The calibration of the phase measuring deflectometry (PMD) system is an important factor affecting the accuracy. Aiming at the difficulty of the calibration of posture relationship between the screen and camera in PMD system, a new system calibration method is proposed. By constructing retro-reflectors on the screen, a photogrammetry system is introduced as the medium of transformation between the screen coordinate system and the camera coordinate system. Therefore, a high precision calibration of geometric parameters is achieved. The calibration accuracy is verified using a stereo PMD system. The experimental results show that the reconstruction surface flatness is less than 1 μm, which demonstrates that the proposed method is feasible and effective.

    Tools

    Get Citation

    Copy Citation Text

    Fangming Liu, Jiarui Lin, Yanbiao Sun, Jigui Zhu. Calibration Method for Stereo Phase Measuring Deflectometry System[J]. Laser & Optoelectronics Progress, 2020, 57(5): 051202

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Instrumentation, Measurement and Metrology

    Received: Aug. 14, 2019

    Accepted: Aug. 28, 2019

    Published Online: Mar. 5, 2020

    The Author Email: Zhu Jigui (jiguizhu@tju.edu.cn)

    DOI:10.3788/LOP57.051202

    Topics