Chinese Optics Letters, Volume. 3, Issue 0s, 164(2005)

Analysis of the effects of intraocular lens edge design on dysphotopsia for pseudophakic patients

[in Chinese]1, [in Chinese]1, [in Chinese]1, [in Chinese]2, [in Chinese]2, and [in Chinese]2
Author Affiliations
  • 1School of Optical Engineering, Nanjing University of Science and Technology, Nanjing 210094
  • 2ABLE Eye Device CO. Ltd., Zhengzhou 450001
  • show less

    The dysphotopia that is connected to the reflection of light from the intraocular lens (IOL) edge is analyzed. A new IOL with 'Sigma' edge is investigated, and other three IOLs are also studied for comparison. The interaction of rays as a function of edge design was evaluated in an experimental model of the pseudophakic eye using the ray-tracing program of ZEMAX software (Focus Software, Inc.). The results show the rays crossing the edges of IOLs form an image with the shape of a partial ring or thin arc in the periphery of the retina, the potential of the 'Sigma' edge to lower dysphotopsia for pseudophakic patients is promising.