Journal of Innovative Optical Health Sciences, Volume. 5, Issue 1, 1150002(2012)

THE STUDY ON NOVEL MICROELECTRODE ARRAY CHIPS FOR THE DETECTION OF HEAVY METALS IN WATER POLLUTION

HUI-XIN ZHAO... WEI CAI, DA HA, HAO WAN and PING WANG* |Show fewer author(s)
Author Affiliations
  • Biosensor National Special Laboratory Key Laboratory for Biomedical Engineering of Education Ministry Department of Biomedical Engineering Zhejiang University, Hangzhou 310027, P. R. China
  • show less

    Qualitative and quantitative analysis of trace heavy metals in aqueous environment are rapidly assuming significance along with the rapid development of industry. In this paper, gold microelectrode array (MEA) plated with mercury film was used for simultaneous voltammetric detection of zinc, cadmium, lead and copper ions in water. The electrochemical behavior and the actual surface area of the MEA were investigated by cyclic voltammetry in K3[Fe(CN)6]. Electrochemical impedance spectrum (EIS) was utilized to examine the deposition of mercury on the electrode surface. Based on anodic stripping voltammetry, mercury film-Au MEA was applied to the detection of heavy metals in artificial analyte, where good calibrate linearity was obtained for cadmium, lead and copper ions, but with zinc exhibiting poor linearity.

    Tools

    Get Citation

    Copy Citation Text

    HUI-XIN ZHAO, WEI CAI, DA HA, HAO WAN, PING WANG. THE STUDY ON NOVEL MICROELECTRODE ARRAY CHIPS FOR THE DETECTION OF HEAVY METALS IN WATER POLLUTION[J]. Journal of Innovative Optical Health Sciences, 2012, 5(1): 1150002

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Oct. 2, 2011

    Accepted: --

    Published Online: Jan. 10, 2019

    The Author Email: WANG PING (cnpwang@zju.edu.cn)

    DOI:10.1142/s1793545811500027

    Topics