Acta Optica Sinica, Volume. 38, Issue 4, 0411006(2018)

Research on Response Curve Calibration Method of Imaging System Based on Digital Micromirror Device

Wei Feng1,2, Xinghua Qu1, Weijing Wang1, and Fumin Zhang1、*
Author Affiliations
  • 1 State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072, China
  • 2 School of Mechanical Engineering, Hubei University of Technology, Wuhan 430068, China
  • show less

    High dynamic range imaging technology is a research hotspot in the field of optical imaging. Digital microscopy device (DMD) has the characteristics of modulation incident ray spatial information. A high dynamic range imaging system with high-resolution is developed based on DMD, which could improve the dynamic range of the imaging system. At the same time, in order to reasonably and effectively control the DMD, the response curve of DMD modulation and CMOS image detector are calibrated by using mathematical analysis method, and the response curve calibration method of DMD camera is proposed according to the structure and imaging characteristics of the imaging system. The experimental results show that the imaging system has a dynamic range larger than the ordinary ones, which can reach more than 96 dB. The fast response function calibration of the DMD camera is also completed, which provides the powerful foundation for the adaptive adjustment of the DMD mask.

    Tools

    Get Citation

    Copy Citation Text

    Wei Feng, Xinghua Qu, Weijing Wang, Fumin Zhang. Research on Response Curve Calibration Method of Imaging System Based on Digital Micromirror Device[J]. Acta Optica Sinica, 2018, 38(4): 0411006

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Imaging Systems

    Received: Aug. 31, 2017

    Accepted: --

    Published Online: Jul. 10, 2018

    The Author Email: Zhang Fumin (zhangfumin@tju.edu.cn)

    DOI:10.3788/AOS201838.0411006

    Topics