Acta Optica Sinica, Volume. 31, Issue 10, 1004001(2011)

Method for Displacement Calibration of Four-Quadrant Photodetector in Optical Tweezers System

Huang Lu*, Guo Honglian, Qu E, Ling Lin, Li Zhaolin, Zhang Daozhong, Li Zhiyuan, and Feng Baohua
Author Affiliations
  • [in Chinese]
  • show less

    A method for real-time, quick calibration of four-quadrant photodetector (QD) signals in optical tweezers system is provided. The experimental results indicate that the method is feasible and precise. Its instability is less than 1%. In addition, the effect of the bead depth on calibration factor is investigated experimentally, and it is found that the calibration factor decreases when the bead depth increases. The method can be used to monitor the stability of the optical tweezers system.

    Tools

    Get Citation

    Copy Citation Text

    Huang Lu, Guo Honglian, Qu E, Ling Lin, Li Zhaolin, Zhang Daozhong, Li Zhiyuan, Feng Baohua. Method for Displacement Calibration of Four-Quadrant Photodetector in Optical Tweezers System[J]. Acta Optica Sinica, 2011, 31(10): 1004001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Detectors

    Received: Mar. 21, 2011

    Accepted: --

    Published Online: Sep. 29, 2011

    The Author Email: Lu Huang (huanglu@aphy.iphy.ac.cn)

    DOI:10.3788/aos201131.1004001

    Topics