Acta Optica Sinica, Volume. 32, Issue 3, 322007(2012)

Design Method of Optical System of Improved Czerny-Turner Imaging Spectrometer

Liu Jianpeng*, Tang Yi, Huang Gang, Zhang Zhige, and Ni Guoqiang
Author Affiliations
  • [in Chinese]
  • show less

    Astigmatism is the main aberration that limits the spatial resolution of Czerny-Turner imaging spectrometer. Focal power is introduced to calculate astigmatism, and formulas are deduced to correct it. The method of computing the distance between grating and collimator is given, which is able to correct the aberration of the edge of field of view. The computing method of tilt angle of imaging plane, which is helpful to correct the aberration at all wavelengths. And an imaging spectrometer based on such structure working in 115~200 nm is designed, with f′=48 mm, F number is 5.0 and image size is 8 mm×7 mm. The designed imaging spectrometer can get spectral resolution δλ=0.22 nm and modulation transimission function (MTF) is 0.7 at all wavelengths and full field of view. The design method can be applied to other imaging spectrometers of different structures.

    Tools

    Get Citation

    Copy Citation Text

    Liu Jianpeng, Tang Yi, Huang Gang, Zhang Zhige, Ni Guoqiang. Design Method of Optical System of Improved Czerny-Turner Imaging Spectrometer[J]. Acta Optica Sinica, 2012, 32(3): 322007

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: Optical Design and Fabrication

    Received: Aug. 29, 2011

    Accepted: --

    Published Online: Mar. 6, 2012

    The Author Email: Jianpeng Liu (liujp010@126.com)

    DOI:10.3788/aos201232.0322007

    Topics