Acta Photonica Sinica, Volume. 48, Issue 6, 630001(2019)

LIBS Autofocus Method Based on Acutance of Surface Image

WANG Hua-dong1,2、*, FU Hong-bo1, JIA Jun-wei1,2, ZHOU Qi-qi1,2, NI Zhi-bo1, and DONG Feng-zhong1,2
Author Affiliations
  • 1[in Chinese]
  • 2[in Chinese]
  • show less

    By using a homemade laser induced breakdown spectroscopy system integrated with a micro-imager, the influence of aperture, region of interest, reflectivity of sample on position sensitivity of image acutance was analyzed. The results show that image acutance is more sensitive to the position under larger aperture, larger aperture is more suitable for the autofocus process. The reflectivity of sample has a little influence on the position sensitivity of image acutance, and the acutance of the partial image can also be used to assist the laser induced breakdown spectroscopy focus. The surface position corresponding to the maximum acutance has good repeatability for different samples and different region of interest. On this basis, an autofocus method based on the acutance of the surface image was established.

    Tools

    Get Citation

    Copy Citation Text

    WANG Hua-dong, FU Hong-bo, JIA Jun-wei, ZHOU Qi-qi, NI Zhi-bo, DONG Feng-zhong. LIBS Autofocus Method Based on Acutance of Surface Image[J]. Acta Photonica Sinica, 2019, 48(6): 630001

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Received: Jan. 15, 2019

    Accepted: --

    Published Online: Jul. 10, 2019

    The Author Email: Hua-dong WANG (whdm818@163.com)

    DOI:10.3788/gzxb20194806.0630001