Acta Optica Sinica, Volume. 25, Issue 10, 1429(2005)

Study on the Response Characteristic of CsI∶Tl Crystal to High-Energy X-Ray Radiation

[in Chinese]*, [in Chinese], [in Chinese], and [in Chinese]
Author Affiliations
  • [in Chinese]
  • show less

    High efficiency CsI∶Tl crystal is used as X-ray converter in the detection system. The response characteristic of CsI∶Tl crystal to X-ray is very important in precision image processing. It is difficult to obtain a stabile H-D curve of X-ray film because of some uncontrollable factors in its processing. The response of CsI∶Tl crystal to X-ray has been theoretically proved to be linear. Some experiments on the 60Co X-ray source are performed to prove the linear of the response and the quantificational H-D curves of the CsI∶Tl crystal with different width are obtained. These H-D curves definitely show that the response is linear and that the linearity is very good. The linearity of less 0.15% is obtained from the experiment. This solves some unconquerable problems such as obtaining a stable, repetitive H-D curve which have existed for a long time in a X-ray film system.

    Tools

    Get Citation

    Copy Citation Text

    [in Chinese], [in Chinese], [in Chinese], [in Chinese]. Study on the Response Characteristic of CsI∶Tl Crystal to High-Energy X-Ray Radiation[J]. Acta Optica Sinica, 2005, 25(10): 1429

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category: X-Ray Optics

    Received: Nov. 17, 2004

    Accepted: --

    Published Online: May. 23, 2006

    The Author Email: (j_xg_caep@sohu.com)

    DOI:

    Topics