Journal of Atmospheric and Environmental Optics, Volume. 5, Issue 5, 366(2010)

Absolute Bidirectional Reflectance Factor Measurement for Diffuse Reflection Standard

Hong-yao CHEN*, Li-ming ZHANG, Ben-yong YANG, and Wei-wei XU
Author Affiliations
  • [in Chinese]
  • show less

    A gonioreflectometer which concentrated on absolute measurement of diffuse reflection standards’ bidirectional reflectance factor (BRF) is described. The instrument employs a specially built integrating sphere as a light source and a mechanically simple, radiometrically stable filter trap radiometer as a detector. Analysis of its initial performance in 630 nm shows that the combined standard uncertainty of the measurement is 0.87 % (k=2). The hemispherical reflectance factor is obtained through integration of BRF measurement results, and it is 0.57% higher than the results scaled in National Institute of Metrology (NIM) using integrating sphere method. The difference is within the uncertainty of the two measurements, and it can validate the accuracy of absolute measurement.

    Tools

    Get Citation

    Copy Citation Text

    CHEN Hong-yao, ZHANG Li-ming, YANG Ben-yong, XU Wei-wei. Absolute Bidirectional Reflectance Factor Measurement for Diffuse Reflection Standard[J]. Journal of Atmospheric and Environmental Optics, 2010, 5(5): 366

    Download Citation

    EndNote(RIS)BibTexPlain Text
    Save article for my favorites
    Paper Information

    Category:

    Received: Mar. 31, 2010

    Accepted: --

    Published Online: Dec. 7, 2010

    The Author Email: Hong-yao CHEN (hyaochen@mail.ustc.edu.cn)

    DOI: